15 

OZL-45-BA-e-1510

5.9  Changing the bulb 

(OZL 451) 

Halogen 

Before  changing  the  halogen  bulb(s),  you  must  always  switch  off  the  device  and 
unplug it from the mains.  You must  also make sure that  the bulb and housing have 
cooled down, so that you avoid any risk of possible burn injuries. 

Please  always  use  cloth  gloves  or  similar  to  hold  and  fit  the  new  bulb  otherwise 
grease and dust residue on the surface of the bulb could have a negative effect on its 
brightness and service life. 

To  change  the  halogen  bulb  of  a 

transmitted  light

  unit,  you  must  first  undo  the 

screw  for  the  bulb  holder  on  the  bottom  of  the  stand,  so  that  you  can 

flip  out  the 

plate

Then you can simply 

pull

 the bulb 

out of its socket

 and replace it with a new one. 

When you have done this, you must make sure that the bulb holder on the underside 
of the stand is re-fitted properly. 

Important:

 When changing the transmitted light bulb, it is helpful if you first 

remove 

the holder and head of the microscope from the pillar

. In this way you can easily 

rotate the stand and replace the bulb. 

To change the halogen bulb of a 

reflected light unit

, you must first unscrew the bulb 

housing with one hand. There is a risk here that the user could be 

burnt

. Even if it is 

only operated for a few minutes, 

the housing will heat up significantly

You  must be  particularly careful when  removing  the  housing, as the 

colour filter

 is 

not fixed on the underside and 

can

 easily 

fall out

, if it is rotated. 

Then you can simply 

pull

 the bulb 

out of its socket

 and replace it with a new one. 

Attention:

 When  selecting  a  replacement  bulb  for  the  reflected  light  unit  you  must 

use  a  halogen  bulb  connected 

to a  screen

 which  should  improve the beam 

characteristics. 

5.10  Changing the fuse 

(OZL 451) 

The  fuse,  or  its  holder  is  fitted  is  on  the  rear  of  microscope  stand  base  below  the 
mains connector. 
If the fuse has blown, then with the device switched off and the power disconnected, 
the fuse can easily be pulled out and replaced with a new fuse. 
You can use a flathead screwdriver to help pull out the holder. 

Summary of Contents for OZL-45

Page 1: ...ohn GmbH Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Stereo zoom microscope KERN OZL 45 OZL 451 OZL 4...

Page 2: ......

Page 3: ...5 Operation and functionality 10 5 1 Getting started 10 5 2 Adjusting the interpupillary distance 10 5 3 Adjusting the magnification 10 5 4 Dioptre adjustment and focussing 11 5 5 Adjusting the stand...

Page 4: ...make sure that you are using the correct input voltage The information to select the correct power supply is located on the device on the rear of the stand base You must comply with this information I...

Page 5: ...older must not be rotated back too far If you do this there is a risk that the microscope could tip over When the microscope is not being used you should fit the objective cap and cover the microscope...

Page 6: ...s these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organ...

Page 7: ...OZL 45 BA e 1510 6 2 Nomenclature Using the OZL 451 as an example with lighting...

Page 8: ...55 75 mm Dioptre adjustment On both sides Packaging dimensions WxDxH 345x320x470 mm Standard configuration Model KERN Tube Eyepiece Field of view mm Objective Zoom Stand Illumination OZL 451 Binocula...

Page 9: ...ighting control for example we recommend that you position the head with the tube connectors centrally to the front Ideally the holder and head are then parallel on the central axis of the stand base...

Page 10: ...9 OZL 45 BA e 1510 Assembled stereo zoom microscope OZL 453...

Page 11: ...nterpupillary distances So each time a different person uses the microscope the gap between the two eyepieces must be re adjusted While you are looking through the eyepieces use one hand to hold the r...

Page 12: ...ugh the right eyepiece with the right eye and bring the object into focus by using the focus control dials 5 Now set the largest possible zoom factor 6 Once again still only looking through the right...

Page 13: ...could occur if the objective lens and the object being observed should collide Height adjustment Fixing the microscope holder With a stereo microscope on a pillar stand the microscope head can be hei...

Page 14: ...ied by the glasses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved W...

Page 15: ...tandard version the lighting is not suitable for the application then it often makes sense to fit an external lighting unit to overcome this problem The lighting units which are suitable for devices o...

Page 16: ...irst remove the holder and head of the microscope from the pillar In this way you can easily rotate the stand and replace the bulb To change the halogen bulb of a reflected light unit you must first u...

Page 17: ...5x 37 5x 255x Field of view mm 15 8 2 4 31 5 4 8 10 5 1 6 7 9 1 2 Working distance 113 mm 220 mm 50 mm 35 mm 7 Features Model outfit Kern model Order number OZL 451 OZL 453 Eyepieces HWF 5x 23 2 mm O...

Page 18: ...sufficient The wrong bulb has been used The input voltage is too low Optical unit Problem Possible causes You can see two images The interpupillary distance is not set correctly The magnifications of...

Page 19: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Reviews: