A15.1018 

Polarizing Microscope, Transmit Light 

 

Instruction Manual 

 
 
 
 
 
 
 
 

 

 

 

To ensure the safety and obtain satisfactory performance, please study this instruction manual thoroughly be

fore start to use the instrument.

 

Summary of Contents for A15.1018

Page 1: ...5 1018 Polarizing Microscope Transmit Light Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument ...

Page 2: ...1 CONTENTS CONTENTS 1 USER NOTIECE 2 1 NOMENCLATURE 4 2 ASSEMBLY 5 2 1 Assembly Diagram 5 2 2 Detailed Assembly Procedure 6 3 USING THE CONTROLS 9 4 POLARIZED LIGHT OBSERVATION 11 5 SPECIFICATIONS 13 ...

Page 3: ...y cool designated bulb 24V 100W halogen lamp 7 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to disassemble them yourself 2 The nosepiece and coarse fine focus unit have a compact and precise frame so please don t disassemble them as possible as you can 3 Keep the instrument clean wipe dust regularly and be attention ...

Page 4: ...the microscope That will affect the function or decline the performance of the microscope 8 Place the instrument in a cool dry position After using the microscope remember to cover it with dust helmet Do wait for the lamp house cooling completely before cover ...

Page 5: ...4 1 NOMENCLATURE Video port Eyepiece Objective Revolving round stage Lamp housing Trinocular viewing head Reflected illumination system Bertrand lens unit Main body ...

Page 6: ... various components The numbers indicate the order of assembly When assembling the components make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces Keeping the hexagon spanner well when change the spare parts you will use it ...

Page 7: ... and the guide board fixed together 2 2 2Mouting Revolving Round Stage 1 Loosen the centering knobs 2 Position the stage with the venire scale in front and carefully lower the stage onto the round dovetail on the stage bracket with the positioning pin on the stage aligned with the groove on the front of the stage bracket then loosely tighten the centering knobs Turn the stage to check it is mounte...

Page 8: ...rstly then replace with higher magnification objective if necessary When replacing the objective slowly turning the nosepiece until you hear clicked that means the objective enter the required position the light path center 2 2 4Mounting reflected illumination system Fig 6 1 Mount the Illumination System into the round dovetail on the head of the microscope body properly and use the Allen screw dr...

Page 9: ...crew down the bolt to fix it For transmitted light illumination the power jack is used For reflected light illumination the power jack is used 2 2 10 Mounting Polarizer and Analyzer 1 Insert the polarizer into its slot with the silkscreen surface toward the front and make it in the light path center 2 push the analyzer into its slot NOTE the polarizer and analyzer are engaged when pushed in to the...

Page 10: ...g Compensator Vernier scale 平台支紧螺钉 Polarizing 调节旋钮 起偏器转环 勃氏镜转盘 Analyzer Bertrand lens unit Bertrand lens focusing plate Compensator clamping knob Polarizer clamping knob Polarizer rotation ring Polarizing ruler adjustment knob Stage rotation clamping knob ...

Page 11: ...op position the specimen is moved 45 to its diagonal position To release the 45 click stop function push back the lever 3 2 3 Using the Bertrand Lens By revolving the Bertrand lens dial the Bertrand lens can be selected At the O position the lens is removed from the light path At the B position the lens is engaged 3 2 4 Focusing the Bertrand Lens During conoscopic observation to focus the conoscop...

Page 12: ...the stage decline slightly due to self weight the tension adjustment collar can be adjusted 4 1 2 Adjusting the optical axis Fig 13 The revolving nosepiece is centerable Before leaving the factory the optical axis of the objective and the center of stage has been adjusted coaxial please do not adjust them unless necessary Make sure the 10X objective is inserted into the primary hole on the centera...

Page 13: ...ge the polarizer and analyzer for extinction position 2 Swing the condenser top lens into the light path 3 revolve the Bertrand lens dial to B position to engae the lens into the light path 4 open the aperture iris diaphragm to largest size 5 revolve the focusing dial of the Bertrand lens to focus on the conoscopic image 3 Moving the specimen only move a new specimen detail into the cross line cen...

Page 14: ... eyepiece 10 10 10 10 10 10 objective 5 10 20 40 50 100 Total magnification 50 100 200 400 500 1000 5 Rotatable stage Diameterф170mm 360 scale Minimum reading 6 by means of vernier scale 6 Coaxial Coarse and Fine Focusing System Sensitivity and Graduation of Fine Focus 0 001mm Focusing movement range 32mm 7 Polarizer can be swung out 8 Analyzer 0 360 rotatable 9 Bertrand lens can be removed from o...

Page 15: ...able quintuple Test plate compensato r Λtest plate first order red 1 4λtest plate Quartz wedge test plate Empty test plate filter ND25 ND6 fitler Blue yellow green frosted illumination 24V100W halogen lamp Spare lamp eyepiece 10 plan eyepiece 2 2 2 10 Crisscross eyepiece 10 eyepiece with scale of crosshair 10 mesh eyepiece Strain free objective Plan 5 Plan 10 Plan 20 Plan 40 Plan 100 oil Plan 50 P...

Page 16: ...15 User manual COC Packing list Package ...