Chapter 1: Setting up the scale
8
The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals
within the scope of quality assurance. For this purpose, the answerable user must define a suitable interval as well as
the nature and scope of this check. Information is available on KERNs home page (
www.kern-sohn.com
) with regard to
the monitoring of balance test substances and the test weights required for this. Test weights and balances can be
adjusted quickly and at a reasonable price in KERNs accredited DKD calibration laboratory (return to national normal).
General:
According to the EU guideline 90/384/EEC balances must be verified officially if they are to be used as follows (legally
regulated area):
a)
For commercial transactions if the price of goods is determined by weighing
b)
For the production of medines in pharmacies as well as for analyses in the medical and pharmaceutical
laboratory
c)
For official purposes
d)
For the production of finished packages
In case of doubt, please contact your local office of weights and measures.
Verification information
An EU qualification approval is available for those balances marked as appropriate for verification in the technical data.
In the event that the balance is applied in an area subject to verification as described above, it must be officially verified
and re-verified at regular intervals.
Re-verification of a balance is carried out in compliance with the respective legal provisions of the states. The term of
verification validity for balances in Germany, for example, is normally 2 years.
The legal provisions of the country of use are to be observed.
1.7
Verification
1.6
Monitoring the test substances