Ö
Weigh the material, the net weight will be indicated.
246
g
The taring process can be repeated any number of times,
e.g. when adding several components for a mixture
(adding). The limit is reached when the whole weighing
range is exhausted.
The weight of the weighing container will be displayed as a
minus number after removing the weighing container.
The tare weight is saved until it is deleted.
Delete tare
Ö
Unload the balance and press the
TARE
button, the zero
display appears.
0
g
PRE-TARE
function
Using this function the weight of a tare vessel is stored. Even
after turning off/on the weighing balance will continue working
with the saved tare value.
Ö
In weighing mode put tare vessel on the weighing plate
Ö
Press repeatedly the
MODE
key until „
PtArE
“ flashing
appears.
Ö
Use
SET
key to store the current weight on the weighing
plate as a PRE-TARE value.
Delete PRE-TARE
value
Ö
Remove all loads from the balance, press
TARE
and press
repeatedly the
MODE
key until „
PtArE
“ flashing appears.
Ö
Use the
SET
key to confirm. The PRE-TARE value is
deleted, the zero display appears.
EMS-BA-e-1112
14