14.1 Error messages
Error message Explication
Remedy
ERR.001
ERR.002
Hardware error
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.005
Memory error
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.100
Heated cover is open for more
than 1 minute during
measuring.
Press
ESC
to abort measuring.
ERR.101
ERR.102
Failure „Temperature sensor“
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.110
Heated cover not closed
properly
Press
ESC
to abort measuring.
TIM.oUT
Measuring start 30 minutes
after resetting to zero
Press
ESC
to abort measuring.
ERR.121
ERR.122
ERR.123
Failure „Heating“
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.124
Measuring lasts too long
Check drying time switch-off
criterion or
∆
M
ERR.200
Failure „Current supply“
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.201
Internal error
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.202
Failure „Electrical voltage“
Switch equipment off and on
again. If the error message
remains inform manufacturer.