43
9 Drying parameters
The equipment can be used to store complete drying programs with drying
parameters individually adapted to the sample.
Four drying modes (Standard drying, Step drying, Slow drying and Rapid drying) and
two switch-off criteria (Time-controlled and Automatic) can be selected.
Further parameters such as drying time, drying temperature and
∆
M (constant weight
loss per 30 seconds) should be set depending on the modes, see table below.
Display
Drying mode
Switch-off
criterion
Drying
time
Drying
temperature
∆
M
AUTO
+
see chap.
9.2.1.1
Standard
Automatic
-
x
x
TIME
+
see chap.
9.2.1.2
Standard
Time-controlled
x
x
-
RAPID
+
see chap.
9.2.1.3
RAPID
Fast drying
Time-controlled
x
x
-
or
Automatic
-
x
x
SLOW
+
see chap.
9.2.1.4
SLOW
Slow drying
Time-controlled
x
x
-
or
Automatic
-
x
x
STEP
+
see chap.
9.2.1.5
STEP
Step drying
Time-controlled
x
x
-
or
Automatic
-
x
x
Select memory location for the drying program prior to selecting the drying
parameters, see chap. 9.1.
Do not exceed the operating time of 1 hour to protect the equipment at
drying temperatures above 180°C.