19
Technical data
Resolution of the display
0,1% water content
0,5°C
temperature
Measuring range
8 to 30 % (60%)
Operation temperature
0°C to 40°C
Temperature measuring range -15°C to 85°C (only measuring head)
Storage temperature
-20°C to 60°C
Temperature compensation
automatically
Power supply
4 pcs. 1,5 Volt AA Alkaline
batteries (for approx. 1000 measurements)
Auto Switch Off
after approx. 6 minutes
Current consumption
55 mA (with light)
Display
128 x 64 matrix display, lighted
Dimensions
740 x 65 x 40 mm
Weight
approx. 450g (incl. batteries)
Degree of protection
IP 40
Scope of supply
device FLK-2,
Rubber
protection
4x1,5Volt AA Alkaline batteries
!IMPORTANT! Please read!
Most common reasons for miss readings
• Product temperature out of application range
Material below 0°C resp. above +40°C may cause faulty measurements. The storage of cold material in a warm storage area usually creates
condensed water which may lead to major measuring errors.
• Discrepancy in temperature between device and material
Please ensure that the device and the material under test are being stored at the same temperature (+/- 3°C) before measuring. Protect your
measuring device from direct sunlight for a reasonable time period before taking a measurement. A high temperature difference has a negative
effect on the stability of the measurement results.
• Wrong calibration curve
Double check the correct selection of the calibration curve before measuring.
• Wet or mouldy material
• Frozen measuring material or material containing snow
This leads to a major decrease in accuracy.
• Direction of plugging in
The Direction of plugging has a great influence on the accuracy. Necessarily follow the instructions according to the beginning of the manual!
• Compressed density outside the application range
If the compressed density differs from that specified, there may be deviations!
• Moving the measuring head after the plug in leads to miss readings!
• Water film at the measuring head
After measuring wet material a water film can arise on the sensor head. This could lead to a too high result in the following measurements.
After measuring wet material clean both black plastic parts of the measuring head accurately with a dry cloth.
ATTENTION: Risk of injury by measuring head! Keep away from children younger than 16 years!