BOP-1KW-GL 111315
B-13
B.44
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]? QUERY
CURR:TRIG?
Syntax:
Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]?
Return Value:
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1
Description:
Returns the current value established by CURR:TRIG command.
B.45
[SOURce:]FUNCtion:MODE COMMAND
FUNC:MODE
Syntax:
Short Form: FUNC:MODE {VOLT | CURR | EXT}
Long Form: [SOURce:]FUNCtion:MODE {VOLT | CURR | EXTernal}
Description:
Establishes the operating mode of the power supply.
VOLT = Constant Voltage mode (CV). CURR
= Constant Current mode (CC). EXT = External. FUNC:MODE VOLT commands power supply to Volt-
age mode, FUNC:MODE CURR commands power supply to Current mode. Commanded mode estab-
lishes parameters (voltage or current) monitored for error conditions. Actual mode depends upon load
conditions. When commanded to Voltage mode, if load conditions cause the power supply to try to
exceed the current limit, the unit will flag an error condition. When commanded to Current mode, if
load conditions cause the power supply to try to exceed the voltage limit, the unit will flag an error con-
dition.
B.46
[SOURce:]FUNCtion:MODE? QUERY
FUNC:MODE?
Syntax:
Short Form: FUNC:MODE?
Long Form: [SOURce:]FUNCtion:MODE?
Return value:
0 (voltage mode) or 1 (current mode)
Description:
Returns the operating mode of the power supply.
0 = Voltage mode. 1 = Current mode.
B.47
[SOURce:]FUNCtion:MODE:TRIGger COMMAND
FUNC:MODE:TRIG
Syntax:
Short Form: FUNC:MODE:TRIG {VOLT | CURR}
Long Form: [SOURce:]FUNCtion:MODE:TRIGger {VOLT | CURR}
Description:
Establishes the operating mode of the power supply when a TRIGger command is sent.
VOLT =
Constant Voltage mode (CV). CURR = Constant Current mode (CC). FUNC:MODE:TRIG VOLT com-
mands power supply to Voltage mode when a TRIGger command is sent, FUNC:MODE:TRIG CURR
commands power supply to Current mode when a TRIGger command is sent.
B.48
[SOURce:]FUNCtion:MODE:TRIGger? QUERY
FUNC:MODE:TRIG?
Syntax:
Short Form: FUNC:MODE:TRIG?
Long Form: [SOURce:]FUNCtion:MODE:TRIGger?
Return value:
0 = voltage mode or 1 = current mode
Description:
Returns the operating mode which the power supply will enter when a trigger command is
sent.
VOLT = Constant Voltage mode (CV). CURR = Constant Current mode (CC).
B.49
[SOURce:]LIST:CLEar COMMAND
LIST:CLE
Syntax:
Short Form: LIST:CLE
Long Form: LIST:CLEar>
Description:
Clears all list entries by setting all pointers to 0.
Releases control of the Output Off flag at EXT
PROTECT PORT, pins 5 and 6 (see Table 2-6), from the programmed list. Related Commands: All
LIST commands (See example, Figure B-5.)
B.50
[SOURce:]LIST:COUNt COMMAND
LIST:COUN
Syntax:
Short Form: LIST:COUN<int_value 0 to 255> Long Form: LIST:COUNt> <int_value 0 to 255>
Description:
Establishes how many times the list is executed.
Allows user to establish how many times the list
is executed. For LIST:COUN 0, the unit will execute the sequence indefinitely until a VOLT:MODE
FIXED, command is received. (See example, Figure B-5.)
B.51
[SOURce:]LIST:COUNt? QUERY
LIST:COUN?
Syntax:
Short Form: LIST:COUNT?
Long Form: LIST:COUNt?
Return Value: <int_value>
Description:
Identifies how many times the list will be executed.
Returns value set by LIST:COUN command.
(See example, Figure B-5.)