3-30
BOP-1K-GL 111315
The BOP can be conveniently substituted for a standard BOP that is currently being used with
one of Kepco’s BIT cards by using one of the Compatibility Modes provided (see PAR. 3.6.1 and
3.6.2.
The
SYSTem:SET
(PAR. B.144) command can be used to configure the BOP to operate in a
manner consistent with earlier models of Kepco’s 100W, 200W and 400W BOP power supplies.
3.6.1
BIT 4882 COMPATIBILITY.
To replace a standard low power BOP/BIT 4882 card combination with a 1KW BOP-GL with
minimal reconfiguration of existing test setups, send
SYSTem:SET 4882
followed by
MEMory:UPDate INTERFACE
.
3.6.2
BIT 4886 COMPATIBILITY
To replace a standard low power BOP/BIT 4886 card combination with a 1KW BOP-GL with
minimal reconfiguration of existing test setups, send
SYSTem:SET 4886
.
3.6.3
IEEE 488 (GPIB) BUS PROTOCOL
Table 3-8 defines the interface capabilities of the BOP power supply (Talker/Listener) relative to
the IEEE 488 (GPIB) bus (reference document
ANSI/IEEE Std 488: IEEE Standard Digital Interface
for Programmable Instrumentation
) communicating with a Host Computer—Controller (Talker/Lis-
tener). Tables 3-9 and 3-10 define the messages sent by the BOP, or received by the BOP, via
the IEEE 488 bus in IEEE 488 command mode and IEEE 488 data mode, respectively. These
messages are enabled during the “handshake” cycle, with the BOP power supply operating as
either a Talker or a Listener. (See PAR. 3.6.3.1 to change the GPIB address.
TABLE 3-8. IEEE 488 (GPIB) BUS INTERFACE FUNCTIONS
FUNCTION
SUBSET
SYMBOL
COMMENTS
Source Handshake
SH1
Complete Capability (Interface can receive multiline messages)
Acceptor Handshake
AH1
Complete Capability (Interface can receive multiline messages)
Talker
T6
Basic talker, serial poll, unaddress if MLA (My Listen Address) (one-byte address)
Listener
L4
Basic listener, unaddress if MTA (My Talk Address) (one-byte address).
Service Request
SR1
Complete Capability. The interface sets the SRQ line true if there is an enabled ser-
vice request condition.
Remote/Local
RL1
Complete capability. Interface selects either local or remote information. In local
mode the BOP enables the analog programming inputs. The unit automatically
enters remote mode (digital control) when the first SCPI command is received.
Parallel Poll
PP0
No Capability
Device Clear
DC1
Complete Capability. BOP accepts DCL (Device Clear) and SDC (Selected Device
Clear).
Device Trigger
DT1
Respond to *TRG and <GET> trigger functions.
Controller
C0
No Capability