13
REMARKS
DISPLAY
PRESS
LO CUT= Low cut-off; Lowest rate
value to be recognized. All rate read-
ings below the "cutoff" will assume the
"set lo" value. Key in the desired value
and press RST/ENTER to enter dis-
played value.
NOR= Normalizing (averaging) factor
(00.0 to 99.9); Key in the desired value
and press RST/ENTER to enter dis-
played value. Higher settings provide
more normalizing (averaging) for a
more stable display.
This section tells the unit that the high
& low setting are entered in units per
Minutes, Hours or Seconds. Press the
PRGM key to step through choices.
Press RST/ENTER to enter displayed
choice.
TDEC= Totalizer Decimal; Press the
arrow keys to enter in the desired
totalizer decimal. Press RST/ENTER
to enter displayed choice. Entering a
decimal will add resolution to the total.
(i.e. tdec=#####.#; 100 will be dis-
played as 100.0)
TFACT= Totalizer Factor; This factor
allows you to divide the totalizer by 1,
10, 100, 1000
Press the PRGM key to step to the
desired factor. Press RST/ENTER to
enter displayed choice.
Lo CuT
Press any key to view
or change existing
value
NoR
Press any key to view
or change existing
value
NNiNS, HouRS
or
SECS
tdec
tfact
Press any key to view
or change existing
value
1
,
10
,
100
or
1000
STEP
4
CONT...
SETTING
SETUP
ENTER
ENTER
ENTER
ENTER
PRGM
ENTER
ENTER
PRGM