![Kent Scientific SomnoSuite User Manual Download Page 51](http://html1.mh-extra.com/html/kent-scientific/somnosuite/somnosuite_user-manual_1955308051.webp)
47
10/2018 V2.0
SomnoSuite recommended Flow Rates
SomnoSuite Serial numbers Ending in –B
Body Weight
Low-Profile Nose Cone
(mL/min)
Standard Face Mask
(mL/min)
Stereotaxic Mask
(mL/min)
Adjustment
D
DEEFFA
AU
UL
LT
T
M
MA
AN
NU
UA
AL
L
M
MA
AN
NU
UA
AL
L
>15g
10-25
25-30
30-50
15-25g
25-35
30-60
50-70
25-35g
35-50
50-80
70-90
35-45g
44-55
60-90
90-110
45-60g
55-70
80-120
110-140
60-80g
70-90
90-150
140-180
80g-100g
90-100
120-180
180-210
100-120g
100-120
140-200
200-240
120-140g
120-130
150-220
240-270
140-160g
130-150
180-250
270-300
160-180g
150-160
200-270
300-330
180g-200g
160-180
220-300
330-350
200-300g
180-250
230-400
350-480
300g-400g
250-300
320-500
480-580
400-500g
300-360
400-600
600-720
500-600g
360-420
480-680
720-830
600-700g
420-460
550-770
830-920
700-800g
460-520
620-850
920-1000
800-900g
520-560
680-740
1000
, with caution
900-1000g
560-610
740-1000
, with caution
1000
, with caution
1000-1100g
610-660
820-1000
, with caution
Not Recommended
1100-1250
660-730
880-1000
, with caution
Not Recommended
Refer to your User’s Manual for complete instructions.
The SomnoSuite uses the Body Weight setting to adjust the flow rate for a Low Profile Face Mask. If another mask
style is used, adjust the flow rates manually by navigating to
More Vaporizer > Nose Cone Min Vol.
10/2018 V2.0