![Kent Scientific SomnoSuite User Manual Download Page 19](http://html1.mh-extra.com/html/kent-scientific/somnosuite/somnosuite_user-manual_1955308019.webp)
15
Alarms
SomnoSuite alarms alert you to certain conditions that might require attention while
SomnoSuite is running.
There are two levels of alarms:
Level
Based On
Banner
Sound (If On)
Configurable
Warning
Parameters
or other
operating
conditions
Yellow
1-beep pattern
Yes
Serious
Parameters
or other
operating
conditions
Red
3-beep pattern
Yes except for
system alarms
To configure a new Parameter alarm:
1)
From any Run Screen, touch Menu then Config.
2)
Touch Data Mgmt, then Alarms.
i
Silence all Alarms
If you do not wish for the SomnoSuite to make an audible tone when
an alarm is triggered, you can silence them. Turn Dial B to choose
“Silence all Alarms” or “Allow Sound”. Touch Back when finished.
ii
Parameter 1-4
Set a custom alarm based on any measured parameter. You may
choose up to 4. Press A when finished configuring the following
settings:
• Parameters: Choose a parameter to monitor by alarm
• Action: Choose how the alarm will alert you
(message and/or sound)
• Type: Choose status to trigger alarm
• Value: Set the threshold value
• Delay: Choose how long the value must meet the threshold
before triggering the alarm
• Clear: Choose how the alarm needs to be cleared