Zero Sensors
Zeroing the sensors ensures that the internal pressure and flow sensors are
working properly, and calibrates them prior to the ventilation. It is important
that the sensors are zeroed to ensure accurate inspiratory pressure delivery.
1) Set up tubing.
2) From the Vent Run Screen, touch
Setup
then
More Ventilator
.
3)
Touch
Calib & Tests
, then
Zero Sensors
.
4) Press Dial B. When asked to confirm measurement, press Dial B again.
Procedure for Use
RoVent With SomnoSuite Anesthetic Delivery:
1. Place the animal in the induction chamber.
2. From the Anest Run Screen, enter all Setup settings. (See page 8.)
3. From the Vent Run Screen, touch Setup.
1) Enter the ventilator settings.
Tip: The RoVent can ventilate based on a target Inspiratory
Pressure, or by Tidal Volume. The Body Weight Setting will
automatically calculate appropriate Respiratory Rate and tidal
volume settings, dependent on your Priority Setting.
See page 23 for more Advanced Options in Vent Setup.
4. From the Anest Run Screen, touch Induction.
The Run Screen appears and airflow begins automatically at a preset flow rate
of 500mL/min.