12. At the end of the procedure, set the anesthetic concentration to 0%.
13. Touch Ventilator again to stop air flow. Allow the animal to recover.
RoVent without SomnoSuite Anesthetic Delivery:
RoVent can ventilate an animal independent of anesthetic delivery, for cases where you
will use injectable anesthesia, as an example.
1. Use clips to direct airflow to the endotracheal tube only.
SomnoSuite calculates the respiration rate and tidal volume
using the animal’s body weight.
To change these settings, turn Dial A or Dial B. Touch
Setup from the Vent Run Screen to make other changes to
Respiratory Rate, Target Pressure, Tidal Volume, or other
advanced settings.
CAUTION
If you choose to alter the tubing setup, be sure to recalibrate
the deadspace. See page 19.
2. Touch Vent to open the RoVent Run Screen.
3. Touch Start Vent to begin ventilation.
CAUTION
Do not start the ventilator before the animal is intubated and
connected to the ventilator tubing.
4. Enter the Anest Run Screen, and turn the Anesthesia Concentration Knob
to the desired concentration.
5. From the Vent Run Screen, touch Stop Vent to stop air flow.