background image

EN-10

Cleaning

The Controller cabinet can be cleaned with a soft cloth dampened with water.  If necessary, a mild disinfectant and/or detergent can be used; excess 

fluid should be avoided.  The Controller should be wiped with a clean, dry cloth afterward. Do not immerse in any liquid.  Do not use products 

containing ammonium chloride, acetone or other aromatic solvents, as those chemicals will degrade the integrity of the case and cause it to become 

embrittled and possibly crack.

The SCD EXPRESS Compression System cannot be effectively sterilized by liquid immersion, autoclaving, or ETO sterilization, as irreparable damage to 

the System will occur.

Electrical/Electronics Description

Line voltage is fed into the controller through the power cord to the power supply mounted on the Power Supply PC Board in the rear case of the 

controller.  It is important to disconnect the power cord at the outlet before opening the controller case.  Exposure to high voltage on the Power Supply 

PC Board is likely to occur if it is electrically live.  

The power supply converts AC line voltage, 100 to 240 VAC, to DC voltage to power the controller components, including the main Controller PC Board 

that is mounted onto the front case.  Alternately, the main Controller PC Board may be directly powered by the optional battery pack.  The Controller PC 

Board controls all functionality of the system and includes the transducer and buzzer.  It does not contain any high voltage.  The buttons and indicator 

LED’s on the front display of the controller are integrated into the membrane panel which connects to the Controller PC Board.

Kendall does not recommend any attempt to repair printed circuit boards.  In manufacturing, extensive testing is performed that cannot be duplicated in 

the field without specialized equipment.  Improper repair could result in patient or user hazards.

Pneumatic Operation Description

When the controller is turned on, the compressor operates and the valves are cycled to determine which type of garment is connected.  After garment 

detection has completed, an inflation cycle is initiated, releasing air through the set of valves, mounted to a manifold.   A transducer monitors the 

pressure in the garments.  The reading from the transducer assists the controller in adjusting the pump’s motor speed to deliver the proper pressure to 

the garments in the appropriate amount of time.

Section V - Test Methods and Calibration

The SCD EXPRESS Compression System has various test modes that can be accessed by the service technician.  They are intended for use by qualified 

personnel.  To activate the test modes follow these steps for entering “Test Access Mode”.  

1.  Plug the controller into an outlet supplying the appropriate line voltage.  Do not activate the test modes while operating on battery power.

2.  Press the Port A and Port B configuration buttons at the same time while turning the controller on.

3.  The buzzer will beep and the Test Mode 01 indicator will illuminate signifying “Test Mode 01”.

NOTE:

  

The Error Code Indicators 1-8 double as Test Mode Indicators 1-8.

  

4.  The user can cycle through the test modes by pressing the Port A button.  Each test mode is indicated by the illumination of the corresponding test 

mode indicator number.  Pressing the Port A button with the last test mode number illuminated (Test Mode 08) will cycle the test mode back to Test 

Mode 01.

5.  After selecting the desired test mode, the Port B button can be pressed to initiate the test. 

6.  If test access is entered but no test mode is selected within 2 minutes, it is assumed that the test access mode was entered inadvertently and a Low 

Pressure alarm will be triggered.

7.  If a test mode is entered and left idle for 5 minutes the unit will revert back to test access mode selection.

8.  To exit Test Access Mode, turn the controller off.

Test Mode 01 - Burn-In

NOTE: 

Burn-In mode is used in manufacturing to ensure proper assembly, to identify premature failures and to calibrate the battery pack.  

This mode is not generally used outside of the manufacturing environment.

1.  Verify nothing is plugged into the ports on the back of the controller and enter test access mode.  Select Test Access Mode 01. 

2.  Press the Port B button to begin Burn-In.  The compressor will operate and the valves will actuate, releasing air out of the ports.  The process repeats 

continuously until the Burn-In period is complete (approximately 16 hours).

3.  The battery, if equipped, will be discharged then charged for calibration during Burn-in. 

4.  When 16 hours of Burn-In is completed the controller will go into alarm mode, flashing Test Access Mode 01 red, and the Port A Leg , Port A Foot, 

Port B Leg, Port B Foot indicators green.  The buzzer will not pip during this alarm.

Summary of Contents for SCD EXPRESS

Page 1: ...ingVoor Bediening En Onderhoud SISTEMA DE COMPRESSÃO Manual De Funcionamento E Assistência KOMPRESSIOJÄRJESTELMÄ Käyttö Ja Huolto opas KOMPRESSIONSSYSTEM Betjenings Og Servicehåndbog Operation and Service Manual ΣΎΣΤΗΜΑ ΣΥΜΠΊΕΣΗΣ Εγχειρίδιο Λειτουργίας Και Σέρβις TLAKOVÝ SYSTÉM Provozní A Servisní Příručka KOMPRESSZIÓS RENDSZER Kezelési És Karbantartási Kézikönyv СИСТЕМА ПНЕВМАТИЧЕСКОЙ КОМПРЕССИИ ...

Page 2: ...tus 5 Charging the Battery 5 BatteryWarnings 6 Section III Fault Conditions andTroubleshooting 6 Watchdog Circuit 6 Section IV Service and Maintenance 8 Introduction 8 Warranty and Factory Service 8 Fan Screen andVentilation 9 Fuses 9 Electrical Safety 9 Suggested Preventative Maintenance Schedule 9 Cleaning 10 Electrical Electronics Description 10 Pneumatic Operation Description 10 SectionV Test ...

Page 3: ...anifold and Bed Hook Removal Installation 13 Muffler Removal Installation 13 Power Supply Board Removal Installation 14 Fan and Screen Removal Installation 14 Main Controller Board Removal Installation 14 SectionVII Specifications 15 SectionVIII Schematics 16 Figure 1 Parts Assembly Diagram ExplodedView 16 Controller Parts List 16 Figure 2 Pneumatic Electrical Schematic 18 Section IX Pump Diagrams...

Page 4: ...iate postoperative c gangrene or d recent skin graft 2 Severe arteriosclerosis or other ischemic vascular disease 3 Massive edema of the legs or pulmonary edema from congestive heart failure 4 Extreme deformity of the leg 5 Suspected pre existing deep venous thrombosis Foot Compression The SCD EXPRESS Compression System may not be recommended for use with Foot Cuffs on patients with the following ...

Page 5: ...tection occurring while illuminated CE Mark 0123 Front Panel Display 1 4 8 7 6 3 10 11 12 Item Explanation Item Explanation 5 9 13 14 2 1 AC Power Indicator 2 Battery Charging Indicator 3 Battery Status Indicators 1 3 4 Test Mode Error Code Indicators 1 8 5 Power On Indicator 6 Power On Standby Button 7 Port B Foot Indicator 8 Port B Garment Configuration Button 9 Port B Leg Indicator 10 Service Re...

Page 6: ... will be lit to signify foot compression The port configuration buttons must be pressed for each port that is connected to a foot cuff to turn on the corresponding Foot indicator s NOTE Leg sleeve compression is the default garment configuration when the Controller is first powered on Therefore the configuration buttons do not have to be pressed to begin compression therapy when leg sleeves are be...

Page 7: ... then the longer of the two measurements will be used to adjust the time between cycles Garment Compatibility The SCD EXPRESS Compression System is designed for use with KENDALL garment Reorder Codes 9545 Leg Sleeve SmallThigh Length 9545T Leg Sleeve SmallThigh LengthTear Away 9530 Leg Sleeve MediumThigh Length 9530T Leg Sleeve MediumThigh LengthTear Away 9780 Leg Sleeve LargeThigh Length 9780T Le...

Page 8: ...Green Off Off 15 40 minutes left Amber Flashing Off Off 15 40 minutes left Amber Flashing Off Off 15 40 minutes left Amber Flashing Off Off 15 40 minutes left Amber Flashing Off Off 15 minutes left Red Flashing Off Off 15 minutes left Red Flashing Off Off 15 minutes left Red Flashing Off Off 15 minutes left Red Flashing Off Off UnitnotpluggedinandPoweredOn OperatingonBattery UnitPoweredOff chargin...

Page 9: ... poison control center Do not open battery dispose of in fire or short circuit Doing so may cause the battery to ignite explode leak or become hot and cause personal injury Dispose of improperly working or damaged battery packs according to local regulations Charge only with specified chargers according to Kendall s instructions Section III Fault Conditions and Troubleshooting When the microproces...

Page 10: ...n AC power outlet If the problem persists recalibrate or replace the battery pack The battery can be calibrated by using test mode 01 Burn In SectionV System Pressure Valve Error Software Error Pump Error Vent Error Low Battery Alarm Leg Sleeve pressure is not between 35 and 55 mmHg for 12 consecutive cycles Foot Cuff pressure is not between 110 and 150 mmHg for for f 12 consecutive cycles If a va...

Page 11: ...re on a circuit board is suspected the unit should be returned for service It is recommended that the system be returned with the circuit board in place as removal of the board s involves additional risk of mechanical damage and damage from electrostatic discharge ESD Warranty and Factory Service Tyco Healthcare Group LP warrants that your SCD EXPRESS Compression System is free from defective mate...

Page 12: ...ower before any disassembly A potential SHOCK HAZARD exists when the front cover is removed even with the unit turned off NOTE The power supply cord plug serves as the electrical supply mains disconnect device The system should be periodically inspected to ensure its electrical safety Check the resistance of the power cord ground from the plug to the IEC 60320 connector This resistance should not ...

Page 13: ... controller in adjusting the pump s motor speed to deliver the proper pressure to the garments in the appropriate amount of time Section V Test Methods and Calibration The SCD EXPRESS Compression System has various test modes that can be accessed by the service technician They are intended for use by qualified personnel To activate the test modes follow these steps for entering Test Access Mode 1 ...

Page 14: ...t Mode 06 indicator is lit red to indicate that the unit is ready for the 45 mmHg calibration After 45 mmHg 0 1 mmHg of air pressure is applied to the transducer press the Port B button to complete the calibration for this range 8 Test Mode 07 indicator is lit red to indicate that the unit is ready for the 130 mmHg calibration After 130 mmHg 0 1 mmHg of air pressure is applied to the transducer pr...

Page 15: ...while reading the pressure at 20 seconds and 40 seconds The alarm will pip at the start of the leak test and at the 20 second and 40 second intervals 7 If there is a significant pressure drop over that 20 second time period a Low Pressure alarm will be triggered 8 If no leak is detected the unit will repeat the leak test for Port B Valves 4 6 9 If no leaks are detected at either port then the syst...

Page 16: ... case Disconnect the compressor output tube at the check valve Diagram 02 Remove the compressor intake tubing from the muffler port Diagram 02 Remove the two 2 screws from the compressor mounting bracket Screws are accessible through the right side battery compartment once the battery pack is removed Diagram 01 Slide the entire compressor assembly out of the rear case housing Installation is the r...

Page 17: ...pin fan connector from the power board Observe the angle and orientation of the fan and screen prior to removal to ease reassembly Diagram 04 Slide the fan and screen out of the case by pulling straight out Installation is the reverse of removal For optimum cooling and quietness use only Kendall replacement fans Main Controller Board Removal Installation CAUTION Use a grounded strap when handling ...

Page 18: ...al Bed Hook Yes Optional Bed Hook Power Cord Storage Yes Audible Visual Alarms Low Pressure High Pressure Internal Electronics Malfunction Power Cord Hospital Grade Plug Controller Dimensions Height 6 2 inches 15 8 cm Width 7 0 inches 17 8 cm Depth 4 5 inches 11 4 cm ControllerWeight 3 5 lbs 1 6 kg Power Requirements 100 240VAC 50VA 50 60 Hz Battery 16 4V 2000mAhr Lithium Ion Optional RunTime 6 8 ...

Page 19: ...t Enclosure 3 Main Controller Board 4 Valve Manifold Assembly 5 Wire Barrier 6 Compressor Assembly 7 Power Supply Board 8 Fuses Set of 2 9 Muffler 10 Fan 11 Fan Screen 12 Rear Enclosure 13 Rubber Feet Set of 2 14 Battery Pack 15 Power Cord 16 Cord Capture Door 17 Bed Hook Controller Parts List 17 16 15 13 3 ...

Page 20: ...EN 17 14 12 11 10 9 8 7 5 4 6 2 1 Section VIII Schematics Figure 1 Parts Assembly Diagram ExplodedView ...

Page 21: ...EN 18 Section VIII Schematics Figure 2 Pneumatic Electrical Schematic ...

Page 22: ...EN 19 Section IX Pump Diagrams Diagrams 01 02 03 04 Diagram 01 Compressor Mounting Bracket Screws Valve Manifold Bed Hook Screws Diagram 03 TransducerTube ...

Page 23: ...EN 20 Diagram 04 Diagram 02 Muffler IntakeTube Compressor IntakeTube Compressor OutputTube Tube Section IX Pump Diagrams Diagrams 01 02 03 04 ...

Page 24: ...1 1 2 IEC 61000 4 5 2001 Complies EN60601 1 2 IEC 61000 4 5 2001 Complies Voltage Fluctuations Flicker Voltage Fluctuations Flicker EN61000 3 2 A1 2001 EN61000 3 2 A1 2001 EN61000 3 3 A1 2001 Complies EN61000 3 3 A1 2001 Complies Harmonic Current Emissions Harmonic Current Emissions EN 61000 3 2 2001 EN 61000 3 2 2001 EN61000 3 2 2001 Complies EN61000 3 2 2001 Complies establishments and those dir...

Reviews: