Alarm Code
10
Battery Error
(Battery Indicator 1 )
11
(Error Indicators 5 & 7)
Fault Type
Battery Error
Temperature
Error
Description
If a battery calibration error or battery cell
failure is detected this error will be displayed.
If the internal case temperature of the
Controller drops below 5° C
(41° F) or exceeds 55° C (131° F).
Troubleshooting
- Turn the Controller off, plug it into an
Turn the Controller off
Turn the Controller off
AC power outlet, and restart.
- If the problem persists recalibrate or
replace the battery pack.
- Make sure the controller
is not covered and that the fan port
is not obstructed.
- Turn the Controller off, allow it to
cool, and restart.
- Verify that the cooling fan is
properly connected.
- Have the Controller serviced by a
professional if the problem persists.
Section IV - Service and Maintenance
This service manual is intended for use as a guide to technically-qualified personnel when evaluating System malfunctions. It is not to be construed as
authorization to perform warranty repairs. Unauthorized service will void the warranty.
Introduction
The SCD EXPRESS Controller contains no user serviceable parts. User maintenance is covered in the sections that follow. All other maintenance must be
performed by appropriately qualified service technical personnel.
Service technicians should be familiar with the operator’s portion of this manual and the operating principles of the SCD EXPRESS Compression System.
If a Controller is to be returned to Kendall for service, a description of the operating conditions and the fault code displayed should accompany the unit.
The fault codes displayed by the Controller are useful in diagnosing service problems.
This manual describes service procedures to the board level, with an exploded view of the Controller shown in
Figure1. If a component failure on a circuit board is suspected, the unit should be returned for service. It is recommended that the system be returned
with the circuit board in place, as removal of the board(s) involves additional risk of mechanical damage and damage from electrostatic discharge (ESD).
Warranty and Factory Service
Tyco Healthcare Group LP warrants that your SCD EXPRESS Compression System is free from defective material and workmanship. Our obligation
under this warranty is limited to the repair of Controllers returned to a service center, transportation charges prepaid, within one year of delivery to the
original purchaser. Specifically, we agree to service and/or adjust any Controller as required if returned for that purpose, and to replace and repair any
part which, upon our examination, is proven to have been defective. This warranty does not apply to the Tubing Set or the disposable garments, or
to equipment damaged through shipping, tampering, negligence, or misuse, including liquid immersion, autoclaving, ETO sterilization, or the use of
unapproved cleaning solutions.
To the extent permitted by applicable law, this limited warranty does not cover, and is intended to exclude, any and all liability on the part of the
TT
Company, whether under this limited warranty or any warranty implied by law, for any indirect or consequential damages for breach hereof or thereof.
Except as expressly provided above in the limited warranty, to the extent permitted by applicable law, the Company hereby negates and disclaims
all express and to the extent permitted by applicable law, implied warranties, including the warranties or merchantability and fitness for a particular
purpose.
Controllers requiring repairs should be sent to a service center. Call one of the service centers listed. Obtain a return material authorization number and
ship the controller, prepaid and insured in the original carton.