Section/page:16.5
Parameters
KEN DW 412
Subject to alterations
17/9/2013
Detergent adjustments
Description
Adjustments
from factory
Possible
adjustments
User adjustments
Detergent dosing
Conductivity
Conductivity
Time control
External
Pre-dosing time
Indicates (in seconds) how long the
detergent pump must run when the
machine is filled with water. This only
applies to conductivity dosing.
30
0-99999
Detergent dosing alarm sensitivity (%)
Sensitivity for when an alarm will go if the
conductivity decreases during time
controlled dosing. Decreasing of
conductivity can be due to an empty
detergent canister.
85
0-100
Time controlled dosing (sec)
Indicates number of seconds the detergent
pump must run at each wash. This only
applies to time controlled dosing.
0
0-999
Detergent increasement (%)
This parameter affects the conductivity
based dosing. When calibration of
detergent is done, the dosing can be
increased or decreased by setting this
parameter higher or lower than 100%.
Please note that an increasement of
detergent at 110% does not mean that the
use of detergent is 10% higher than normal.
The conductivity of the detergent is not
proportional with the amount of detergent.
100
0-999