8. Technical Information
46
8-3. Specifications
Model
RA-620
RA-600
Measurement Method
Detection of Critical Angle of Optical Refraction
Light Source
LED Na-D Line (589.3nm)
Measurement Items
Refractive Index, Brix, Other Concentrations
Measurement
Range
Refractive Index (nD)
1.32000 ~ 1.58000
1.3200 ~ 1.7000
Brix
0.00 ~ 100.00%
Accuracy
*1
Refractive Index (nD)
±0.00002
±0.0001
Brix
±0.014%
*2
(0 ~ 85.0%) ±0.1%
Repeatability
*3
Refractive Index (nD)
±0.00001
±0.0001
Brix
±
0.007%(<5%)
±
0.01%(
≧
5%)
±0.1%
Resolution
Refractive Index (nD)
0.00001
0.0001
Brix
0.001% (<5%)
0.01% (
≧
5%)
0.1%
Temperature Control
*4*5
5°C ~ 75°C
5°C ~ 100°C
*6*7
Temperature Indication Resolution
0.01°C
0.1°C
Minimum Amount of Sample
0.2mL
Display
4.7-inch colour TFT LCD
Operation
Touchscreen (Comes with Stylus.)
Security
Password Protection
Data Storage
Number of Methods
100 methods
Measurement Results
300 data
Calibration Record
20 data
Check Record
20 data
External Storage
USB Flash Drive
Temp.
Compensation
Brix
5.00 ~ 75.00°C (Automatic compensation by
preprogrammed conversion table.)
Concentration
By Conversion Table
100 data
Interfaces
LAN
:
x 1
Personal computer (PC)
USB1.1
: x 2
USB flash drive, keyboard, barcode reader,
Epson inkjet printer
*8
, Thermal Printer(DP-600)
RS-232C : x 2
Dot Matrix Printer(IDP-100), Multiple Sample Changer
(CHD-502), Auto Clean and Sampling Unit(DCU-551)
Ambient
Conditions
Temperature
5 ~ 35°C
*7
Humidity
85%RH or below (No condensation allowed.)
Power Supply
AC 100 ~ 240V, 50/60Hz (Comes with AC adapter.)
Power Consumption
20W (max. 50W, min. 10W)
Dimensions
192 (W) × 281 (D) × 166 (H) mm
Weight
5kg
Materials in Contact
with Samples
Prism
Artificial Sapphire
Sample Stage
SUS316
Option
Printer
DP-600, IDP-100
Expandability
Barcode Reader
Reads sample name, measurement conditions,
value of standard liquid
PC Software
Data Acquisition Software
*1: By KEM’s standard measurement conditions.
*2: Calculated from measurement accuracy of refractive index: nD 0.00002 = Brix 0.014%.
*3: By KEM’s standard measurement conditions. Subject to sample properties.
*4: Peltier Thermostat *5: Lower limit 12°C below ambient. *6: 75°C to 100°C – Optional
*7: When the set temperature is above 75°C, reduce ambient temperature to 25°C or below. *8: Enquire for applicable models.
Summary of Contents for RA-600