7. Troubleshooting
RA-620/RA-600
37
Other error message
Error Message
Possible Cause
Remedies
Temp error inside instrument
See
operation
manual.
(E-1400)
Temperature
inside
instrument exceeds upper
limit.
* Upper limit
Set temp ≤75: 60°C
Set temp >75: 50°C
Reduce set temperature or
room temperature.
When the set temperature is
above 75°C, reduce ambient
temperature to 25°C or
below.
Measurement record (list and details)
・
Print data window
Error message at time of Measurement or Check
Contents
Window Display
List of Record
Details
(Error Code)
Not enough sample.
-------
E-1
No Measure
No sample or not enough
sample.
Check sample cover.
-------
E-2
No Measure
Sample cover left open.
Finished at limit time.
!
!
Exceeding limit time.
Exceeding meas range.
*
E-3
*
Summary of Contents for RA-600