Operating manual PT 11x, 12x,13x
32
14.1), we only recommend this for advanced users. If you intend to carry
out measurements of molten metal only, it makes sense to maintain the
factory configuration. This will keep those parameters, which will not re-
quire adjustment, hidden.
Command "S" enables the user to enter a short word to designate the
measurement location/task. This text can be accessed from the Main
Menu by pressing "Q".
------------------------------------------------------------------
Submenu LAMBDA 1
------------------------------------------------------------------
L1 range .... 0.0 - 1000.0 C
L1 User calibration ....... off
L1 User def. offset +0.00000
L1 User def. factor +1.00000
A: Set L1 - extended-range
B: Set L1 User-Cal. On/Off
ESC: Back to MAIN-MENU
------------------------------------------------------------------
>CALIBRATION >LAMBDA 1 >
Use command "A" to reconfigure
the pyrometer‘s measuring range. The
range can be adjusted to be either larger or smaller than the factory-
configured range. When customizing the temperature range, make sure
that upper and lower temperatures you enter are compatible with the py-
rometer
’s actual working range.
Use key "B" for direct access to enable adjustments to the CellaPort PT.
If you make a mistake while making the adjustments, simply enter off-
set=0.0 und factor=1.0, or set User Calibration
to „off“.
To recalibrate your CellaPort, you will require a calibration fur-
nace and a reference standard