Operating manual PT 11x, 12x,13x
21
Parameter
Function
Plausibility
(
TSP_
)
Upper threshold for permissible deviation, if
temperature reading > average value
Plausibility
(
TSP~
)
Lower threshold for permissible deviation, if
temperature reading is < average value
If a measuring cycle does not start during the period
T.OUT
the saved
average will be deleted and reinitialized when the next cycle begins.
Parameter
Function
Timeout
(
T.OUT
)
Time cycle for deleting average value (in
minutes)
At the end of the measuring cycle the pyrometer displays the averaged
temperature value (or the invalid reading „
----
“ ).
Cut-off interval [
T.DIS
]
A cut-off interval (time lag) begins after the measuring time has ended.
This cut-off interval must expire before the next measurement can start
with the cycle starting conditions described above.
Parameter
Function
Cut-off interval (
T.DIS
)
The interval between one completed sam-
pling and the start of a new sampling.
Timeout
[
T.OUT
]
If a measuring cycle does not start during the period
T.OUT
, the saved
average will be deleted and reinitialized when the next cycle begins.
Parameter
Function
Timeout
(
T.OUT
):
Time cycle for deleting average value (in
minutes)
Autoreset Funktion [
A.RST
]
Activate auto reset for the ATD function to run cyclically. Limit 1 will then
be ignored.
Normally, for a new measuring cycle to start, the detected temperature
must have dropped below Limit 1. If, however, you wish to measure a
continuous process (continuous material flow), you must activate Auto
Reset. Limit 1 will then be ignored. Sampling will continue (restarting au-
Summary of Contents for ITS CellaPort PT 120
Page 1: ...Operating manual PT 11x 12x 13x Pyrometer CellaPort PT 11x 12x 13x Item No 106 3102 01 2019...
Page 6: ...Operating manual PT 11x 12x 13x...
Page 64: ...Operating manual PT 11x 12x 13x 58 33 Dimensions...
Page 69: ...Operating manual PT 11x 12x 13x 63...
Page 70: ......
Page 71: ......
Page 72: ......