Operating manual PT 11x, 12x,13x
30
>I/O >LED CONTROL >S
Set status LED source:
0: Off
1: Ready-Signal
2: Lambda 1
3: Lambda 1 premax
4: Ambient Temperature
5: Lambda 1 ATD Trigger
6: Lambda 1 ATD tAct
------------------------------------------------------------------
Your choice>
In the submenu "Options" you can program a keylock. If keylock is acti-
vated and you wish to change a parameter setting, enter Code
P100
for full access. If you enter an incorrect code, you can view the configura-
tion but you will not be able to change the setting.
------------------------------------------------------------------
Submenu OPTIONS
------------------------------------------------------------------
Autoprint ......... on (cyclic)
Print cycle time ..... 0.1 s
Protocol address .......... 001
Display ........... temperature
Key lock .................. off
Unit .................. Celsius
A: Set autoprint function
T: Set output cycle time
P: Set protocol-address
D: Set display function
E: Set key lock
F: Set unit Celsius/Fahrenheit
R: Switch off pyrometer
ESC: Back to MAIN-MENU
------------------------------------------------------------------
>I/O >OPTIONS >
17.3.4 Automatic temperature data output
For continuous serial transmission of temperature data, you must acti-
vate automatic temperature data output.
In the submenu “Options”, command
“A"
activates (on) / deactivates
(off) this function.
If ATD is activated, the temperature data will only be transmitted after
successful completion of measurement.
If ATD is deactivated, press command
"T"
to select the cycle time in
which the current temperature data are transmitted through the serial in-
terface.
Summary of Contents for ITS CellaPort PT 120
Page 1: ...Operating manual PT 11x 12x 13x Pyrometer CellaPort PT 11x 12x 13x Item No 106 3102 01 2019...
Page 6: ...Operating manual PT 11x 12x 13x...
Page 64: ...Operating manual PT 11x 12x 13x 58 33 Dimensions...
Page 69: ...Operating manual PT 11x 12x 13x 63...
Page 70: ......
Page 71: ......
Page 72: ......