Operating instructions CellaTemp PR xx
11 Operation
After connecting the supply voltage the pyrometer will be automatically initialized
and will perform a self-diagnosis� After approx� 0�5 sec the sensor is ready to
operate and the instrument runs the signal processing�
11.1 Ambient temperature
The maximum permissible ambient operating temperature for the pyrometer is
65 °C� If the instrument is used in ambient temperatures above 65 °C, it must be
either cooled or shielded from excess radiant heat by means of a deflector plate�
11.2 Device status
Status
value
State
Triggering event
Measures
0
Normal
operation
1
Mainte-
nance
required
–
2
Out of
specifica-
tion
Maximum permissible
internal temperature
exceeded
Cool down the unit
Calibration data incom-
plete
Perform calibration
Supply voltage unstable
Check supply voltage
3
Function
test
Temperature simulation
function active
Check measured value
4
Non re-
coverable
error
EEPROM memory data
corrupt
Replace unit