14
Operating instructions CellaTemp PR xx
10 Parameterizing
Parameters can be set via the IO-Link interface before installation and
commissioning of the unit or during ongoing operation�
Changing the parameters during operation will affect the functioning of
the system� Before parameters are changed, it must be ensured that no
malfunction occurs in the system�
10.1 Damping function
When the target object‘s temperature is erratic, the damping function smoothens
these temperature fluctuations in order to stabilize the measuring signal� The
greater the time constant [dAP], the lower the effect of these fluctuations on the
yielded temperature reading�
1: Output signal without smoothing function
2: Output signal with smoothing function
Parameter
Ldx.
Subindex /
bitOffset
Value range
Slope
dAP
6120
0
0 … 9999
0�1
10.2 Peakhold-Funktion
It might often be desirable to determine the peak temperature during a defined time
period, for example when the objects to be measured move past the pyrometer,
resulting in temperature readings which would appear to be cyclical� In this mode,