Operating instructions CellaTemp PR xx
the displayed temperature reading will not drop between targeted objects� The peak
temperature reading will be held for a preset time period�
The hold time [
PhLd
] can be set from 0�1 to 999�9 sec� The maximum tempera-
ture sampled during the defined hold time will be saved� It makes sense to choose
a hold time which is approximately 1�5 times as long as the cycle of the moving
targets� This avoids temperature drops� Any changes are recognised at once�
1
2
3
4
5
OUT2
[mA]
t
1: Measuring object in front of pyrometer
2: Hold time
3: second internal hold time
4: Measuring readings with peak hold function
5: Measuring readings without peak hold function
Parameter
Ldx.
Subindex /
bitOffset
Value range
Slope
PhLd
6110
0
0 … 9999
0�1
10.3 Reset all parameters to factory settings
Resetting to factory settings is done via system command [Index 2] -> [Wert 130]
After a reset to factory settings configure the emissivity coefficient [EPS1]
again (→ 9 Setup) to obtain exact measuring results.