Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-33
Example 2
printbuffer(1, 6, smua.nvbuffer1.sourcevalues)
Print the sourced value of the first 6 readings
stored in source-measure unit (SMU) A,
buffer 1.
Example output:
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04
Also see
(on page 7-25)
(on page 7-26)
(on page 7-27)
(on page 7-28)
(on page 7-29)
(on page 7-31)
(on page 7-33)
bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Clearing the buffer
See
Details
Not applicable
Usage
statusInformation
=
bufferVar
.statuses[
N
]
statusInformation
The status value when reading
N
of the specified buffer was acquired
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
N
The reading number (
1
to
bufferVar
.n
)