Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-43
DISPlay subsystem
This subsystem contains commands that control the front-panel display.
:DISPlay:CLEar
This command clears the text from the front-panel USER swipe screen.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:DISPlay:CLEar
Example
DISP:CLE
DISP:SCR SWIPE_USER
DISP:USER1:TEXT "Batch A122"
DISP:USER2:TEXT "Test running"
Clear the USER swipe screen and switch to
display the USER swipe screen.
Set the first line to read "Batch A122" and
the second line to display "Test running".
Also see
(on page 6-47)
:DISPlay:<function>:DIGits
This command determines the number of digits that are displayed for measurements on the front panel.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
See
Functions and defaults
Usage
:DISPlay:<function>:DIGits <n>
:DISPlay:<function>:DIGits DEFault
:DISPlay:<function>:DIGits MINimum
:DISPlay:<function>:DIGits MAXimum
:DISPlay:<function>:DIGits?
:DISPlay:<function>:DIGits? DEFault
:DISPlay:<function>:DIGits? MINimum
:DISPlay:<function>:DIGits? MAXimum
<function>
The function to which the setting applies; see
Functions
<n>
3.5 digit resolution:
3
4.5 digit resolution:
4
5.5 digit resolution:
5
6.5 digit resolution:
6
7.5 digit resolution:
7
(not available for digitize functions)