Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-47
:DISPlay:USER<n>:TEXT[:DATA]
This command defines the text that is displayed on the front-panel USER swipe screen.
Type
Affected by
Where saved
Default value
Command only
Power cycle
Not applicable
Not applicable
Usage
:DISPlay:USER<n>:TEXT[:DATA] "<textMessage>"
<n>
The line of the USER swipe screen on which to display text:
•
Top line:
1
•
Bottom line:
2
<textMessage>
String that contains the message; up to 20 characters for
USER1
and
32 characters for
USER2
Details
This command defines text messages for the USER swipe screen.
If you enter too many characters, the instrument displays a warning event and shortens the message
to fit.
Example
DISP:CLE
DISP:SCR SWIPE_USER
DISP:USER1:TEXT "Batch A122"
DISP:USER2:TEXT "Test running"
Clear the USER swipe screen and switch to
display the USER swipe screen.
Set the first line to read "Batch A122" and
the second line to display "Test running".
Also see
(on page 6-46)