Verify measurement operation
The following steps provide a quick way to verify that the
instrument is operating correctly.
To verify measurement operation:
1. Turn the instrument on.
2. On the front panel, press the
HOME
key.
3. On the Functions swipe screen, select
DCV
.
4. Short the connections.
The voltage measurements appear in the MEASURE DC
VOLTAGE area of the home screen and should read
approximately 0 V.
To view the measurements on the graph screen:
1. Press
the
MENU
key.
2. Under Views, select
Graph
.
You can adjust the graph settings using the options in the Data
and Scale tabs.
Summary of Contents for DMM7510