Section 6: Measuring 4-wire resistance with offset compensation
Model DMM7510 7½ Digit Multimeter User's Manual
6-4
DMM7510-900-01 Rev. B / May 2015
Using the front panel
To set up the application from the front panel:
1.
Press the
POWER
button on the front panel to turn on the instrument.
2.
On the FUNCTIONS swipe screen, select
4W
Ω
to select the 4-wire resistance measure function.
3.
Press the
MENU
key.
4.
Under Measure, select
Settings
.
5.
Set Offset Compensation to
On
.
6.
Press the
HOME
key.
The measurement readings are displayed in the top area of the Home screen.
Using SCPI commands
This sequence of SCPI commands measures the resistance of a device or component. The device
under test is connected to the Model DMM7510 and controlled remotely.
You may need to make changes so that this code will run in your programming environment.
Send the following commands for this example application:
Command
Description
*RST
:SENS:FUNC "FRES"
:SENS:FRES:OCOM ON
:SENS:FRES:AZER ON
:SENS:FRES:NPLC 1
:READ?
•
Reset the Model DMM7510
•
Set to measure four-wire resistance
•
Enable offset compensation
•
Enable auto zero
•
Set NPLC to 1
•
Read the resistance value