Model 8011 Connection Instructions
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
9
6. Press
the
CONFIG
key.
7. Press
the
SRC
or
MEAS
key (the sense mode can be set from either the V-SOURCE or the V-MEAS menu).
8. If you pressed the SRC key, select
V-SOURCE > SENSE-MODE
.
9. If you pressed the MEAS key, select
V-MEAS > SENSE-MODE
.
10. Select
4-WIRE
to enable remote sense.
11. Test the device.
Table 2. Connections for two-terminal device test (remote sense)
Cable end
Cable
Cable end
Model 2651A rear panel
CA-552-1 (supplied with the Model 2651A)
Open
LO of CA-552-1
Black CA-560-0
8010-120 blue terminal 4
HI of CA-552-1
Red CA-560-2
8010-120 yellow terminal 6
Model 2651A rear panel
CA-557-1 (supplied with the Model 2651A)
CS-1629-8
Sense LO of CS-1629-8
Black CA-567-0
8010-120 blue terminal 1
Sense HI of CS-1629-8
Red CA-567-2
8010-120 yellow terminal 3
Figure 6. Connections for two-terminal device test (remote sense)
S LO
G
G
G
G
S HI
2-T
erminal DUT
Plug
8010-317 insulator plug
(insert into 3-pin socket)
HI
HIGH CURRENT TEST SOCKET KIT
8011
RATINGS:
40V MAX.
15A DC MAX.
50A PULSE MAX.
From
Model 2651A
CA-567-0
(black)
CA-560-0
(black)
CA-560-2
(red)
CS-1629-8
CA-552-1
(supplied with
Model 2651A)
CA-557-1
(supplied with
Model 2651A)
CA-567-2
(red)
From
Model 2651A
8010-120 TO-247
Device Test Board