Model 8011 Connection Instructions
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
7
General guidelines
The High Current Test Socket Kit provides several types of connections to allow you to make connections to meet
the needs of your application.
In general, when making connections:
•
Remove power from the SourceMeter instrument before making connections.
•
Make connections to the SourceMeter instrument before making connections to the Model 8011 boards.
•
Connect the force leads of the SourceMeter instrument to terminals 4, 5, and 6 of the Model 8011 boards.
Connect the sense leads to terminals 1, 2, and 3.
•
Install the device under test (DUT) after making the SourceMeter instrument connections.
Use-cases are shown below that demonstrate some typical connections.
Use-cases
Two-terminal DUT using a Model 2651A set to local sense
NOTE
These instructions provide general information about connections to the Model 2651A. For detail on the
connections, see the Model 2651A user documentation.
Tools needed
•
Small flat-head screwdriver
Instructions
1. Remove power from the Model 2651A.
2. Make connections as shown in Table 1 and Figure 5.
3. Install the 8010-317 insulator plug into the test socket.
4. Install the 2-terminal DUT into the axial posts labeled 4, 1, 3 and 6 on the 8010-120 board.
5. Return power to the Model 2651A.
6. Press
the
CONFIG
key.
7. Press
the
SRC
or
MEAS
key (the sense mode can be set from either the V-SOURCE or the V-MEAS menu).
8. If you pressed the SRC key, select
V-SOURCE > SENSE-MODE
.
9. If you pressed the MEAS key, select
V-MEAS > SENSE-MODE
.
10. Select
2-WIRE
to enable local sense.
11. Test the device.