Matrix Configuration
2-3
Single-ended switching
In the single-ended switching configuration, the source or
measure instrument is connected to the DUT through a sin-
gle pathway as shown in Figure 2-4.
Differential switching
The differential or floating switching configuration is shown
in Figure 2-5. The advantage of using this configuration is
that the terminals of the source or measure instrument are not
confined to the same matrix crosspoint. Each terminal of the
instrument can be connected to any matrix crosspoint.
Source or
Measure
Row
Columns
7022
H
L
HI
LO
DUT
Optional
Shield
Figure 2-4
Single-ended switching example
Figure 2-5
Differential switching example
HI
LO
Source or
Measure
Rows
Columns
7022
H
L
H
L
DUT
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Summary of Contents for 7022
Page 89: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 90: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 91: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 92: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 93: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 97: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 98: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 104: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 105: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...