5-8
6517B-900-01 Rev. A / Jun 2008
Section 5: Remote Operation
Model 6517B Electrometer User’s Manual
SCPI signal-oriented commands
Signal-oriented commands are used to acquire readings using a set of high-level instructions to
control the measurement process. The two most commonly used commands are
:FETCh?
and
:MEASure?
.
Details on all signal-oriented commands are covered in Section 14 of the Reference Manual.
:FETCh?
This query command is used to request the latest post-processed reading. After sending this
command and addressing the Model 6517B to talk, the reading will be sent to the computer. This
command does not trigger a measurement, it simply requests the last available reading.
:MEASure[:<function>]?
In general, this command places the instrument in the “one-shot” measurement mode (ARM
indicator off), triggers a single measurement, and then performs a
:FETCh?
command.
If you include the measurement function with this command, the instrument will first go to the
specified function and then perform the one-shot measurement and query, as follows:
:MEASure?
Currently selected function
:MEASure:VOLTage[:DC]?
Volts function
:MEASure:CURRent[:DC]?
Amps function
:MEASure:RESistance?
Ohms function
:MEASure:CHARge?
Coulombs function
SCPI subsystem commands
Detailed information on all SCPI commands is covered in Section 14 of the Reference manual. In
this section, only the commands used for basic operation are summarized.
Defaults
:SYSTem
SYSTem
subsystem
:PREset
Set the 6517B to the system preset defaults. Some of the
selected defaults include: volts, continuous triggers, normal
speed (1.0 PLC), and auto range.
:POSetup <name>
Select power on setups:
RST
,
PRESet
, or
SAV0
through
SAV9
.
Instrument will power up to the
*RST
or system present
defaults, or one of the
*SAV0
–
*SAV9
setups.
Fresh readings
[:SENSe[1]]
SENSe
subsystem
:DATA:FRESh?
Requests a new reading. This command will not request the
same reading twice. If a new reading has already been
triggered, this command will wait until the new reading is
available.
:CALCulate[1]
CALCulate
subsystem
:DATA:FRESh?
Requests a new calculated reading. When
CALC1
is enabled,
SENS:DATA:FRES?
is the reading before calculation and
CALC:DATA:FRES?
is the reading after calculation. When
CALC1
is disabled, the readings are the same.