6517B-900-01 Rev. A / Jun 2008
Return to
3-3
Model 6517B Electrometer User’s Manual
Section 3: Basic Measurements
Figure 3-1
Voltage measurements
Guarding
Guarding should be used for high impedance (>10
9
Ω
) voltage measurements and for voltage
measurements when using long input cables. In a high impedance circuit, guarding greatly
reduces leakage currents. When using long input cables, guarding cancels the effects of cable
capacitance that can significantly slow down the measurement response time.
When GUARD is enabled, the INPUT triax connector is reconfigured to apply the guard potential
to the inner shell of the INPUT triax connector. With this configuration, the COMMON banana jack
is used for input low.
shows both the guarded and unguarded configurations for the
INPUT connector.
Perform the following steps to enable or disable guard:
1.
Press
CONFIG
and then
V
to display the volts configuration menu.
2.
Use the and keys to place the cursor (blinking menu item) on
GUARD
and press
ENTER
.
3.
Place the cursor on the desired selection (
OFF
to disable or
ON
to enable) and press
ENTER
.
4.
Use
the
EXIT
key to back out of the menu.
Red (HI)
Black (LO)
6517B
Shield (Optional)
Vs
Measured
Voltage
+
-
Red (HI)
LO
6517B
Safety
Shield
Vs
+
-
Safety
Earth
Ground
Measured
Voltage
Triax
Cable
A. Unguarded (GUARD off)
Triax
Cable
Guard
Black
Green
B. Guarded (GUARD on)
OPTION SLOT
250VDC MAX
MADE IN
U.S.A
LINE RATING
50-60 Hz
100 VA MAX.
LINE
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD, REPLACE FUSE WITH SAME TYPE AND RATING.
1250Vpk
PREAMP
OUT
1000VDC
MAX
FUSE
630mAT
315mAT
100V
120V
220V
240V
COMMON
2V OUT
LO
HI
V SOURCE
TRIGGER LINK
RS-232
DIGITAL I/O
HUMIDITY
TEMP
TYPE K
INTERLOCK
IEEE-488
2VDC MAX
OPTION SLOT
250VDC MAX
MADE IN
U.S.A
LINE RATING
50-60 Hz
100 VA MAX.
LINE
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD, REPLACE FUSE WITH SAME TYPE AND RATING.
1250Vpk
PREAMP
OUT
1000VDC
MAX
FUSE
630mAT
315mAT
100V
120V
220V
240V
COMMON
2V OUT
LO
HI
V SOURCE
TRIGGER LINK
RS-232
DIGITAL I/O
HUMIDITY
TEMP
TYPE K
INTERLOCK
IEEE-488
2VDC MAX
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com