Model 6517B Electrometer User’s Manual
Section 4: Measurement Options
Trigger connections
Connect the Model 6517B to the switch system as shown in
. Detailed information on
triggers is provided in Section 7 of the Reference Manual.
Figure 4-9
Trigger connections using trigger link
Configure external channels
The scanner configuration menu is used to specify the number of channels (external inputs) for the
scan. If a scanner card is installed in the option slot of the Model 6517B, use procedure A. If the
option slot is empty, use procedure B.
Procedure A
Scanner installed in option slot of Model 6517B:
1.
Press
CONFIG
and then
CARD
to display the scanner configuration menu.
2.
Place the cursor on
EXTERNAL
and press
ENTER
.
3.
You will then be prompted to enter the number of external inputs (channels) for the scan.
Use cursor keys
(
and
)
and the
RANGE
keys (
and
)
to change the number (1 to
400). Press
ENTER
to continue.
4.
Press
EXIT
to return to the normal measurement display.
Procedure B
Option slot of Model 6517B empty:
1.
Press
CONFIG
and then
CARD
to display the scanner configuration menu.
2.
You will then be prompted to enter the number of external inputs (channels) for the scan.
Use cursor keys
(
and
)
and the
RANGE
keys (
and
)
to change the number (1 to
400).
3.
Press
ENTER
to return to the normal measurement display.
Perform the scan
Perform the following steps to scan external channels. Note that menu items are selected by
placing the cursor on it and pressing
ENTER
. A parameter value is changed by using the
RANGE
keys (
and
)
. and pressing
ENTER
.
1.
On the Model 6517B, press the
CARD
key. If a scanner card is installed in the option slot of
the Model 6517B, perform steps a and b. If the option slot is empty, proceed to step 2.
a.
Place the cursor on
PERFORM-SCAN
and press
ENTER
to display the scan types
(internal or external).
b.
Place the cursor on
EXTERNAL
and press
ENTER
.
2.
On the Switch System, if not already done, reset the Model 7001/2 as follows:
a.
Press
MENU
.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATI
MADE IN USA
Model 7001 or 7002 Switch System
6517B
Electrometer
OUT
IN
Trigger
Link
Trigger Link Cable (8501)
OPTION SLOT
250VDC MAX
MADE IN
U.S.A
LINE RATING
50-60 Hz
100 VA MAX.
LINE
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD, REPLACE FUSE WITH SAME TYPE AND RATING
.
1250Vpk
PREAMP
OUT
1000VDC
MAX
FUSE
630mAT
315mAT
100V
120V
220V
240V
COMMON
2V OUT
LO
HI
V SOURCE
TRIGGER LINK
RS-232
DIGITAL I/O
HUMIDITY
TEMP
TYPE K
INTERLOCK
IEEE-488
2VDC MAX
Trigger
Link
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com