Keithley 4299-2 Manual Download Page 2

Model 4299-2 Installation Instructions 

  

 

Installation 

The following procedure outlines the installation of the rack mount kit. 

WARNING  

Make sure the instrument that is being installed is in a powered down state with all 
cables unplugged. Failure to install an instrument in a  discharged state may cause an 
electrical shock or death. 

CAUTION  

To prevent damaging heat build-up, and thus ensure specified performance, adhere to 
the following rack-mounting precautions: 

• 

Keep the rear exhaust vent and at least one side vent free of any obstructions. Even 

partial blockage may impair proper cooling. 

• 

DO NOT position a device adjacent to the Model 260x that forces air (heated or 
unheated) into or onto the Model 260x’s cooling vents or surfaces. This additional airflow 
may compromise accuracy performance. 

• 

Allow at least 1U of space at the top and bottom of the Model 260x (see Note below). 

• 

Make sure there is adequate air flow around at least one side of  
the instrument. Adequate air flow ensures proper cooling. Adequate airflow provides air 
temperatures that are within about one inch of the Model 260x surfaces to remain within 
specified limits under all operating conditions. 

• 

DO NOT rack mount high dissipation equipment adjacent to the Model 260x that could 
cause excessive heating to occur. The specified ambient temperature must be 
maintained around the surfaces of the Model 260x to maintain specified accuracies. 

NOTE  

1U is a standard vertical spacing and is equal to 1.75 inches. Typical distance between the 
mounting holes on rack rails is 0.125” (1/8”). Add vented panel S500-304 (not included) above 
units and Blanking Panel (PNL-10x) below when installing 26xx. 

When determining placement within any rack, make sure to adhere to the Caution above. If the rack has only 

convection cooling, use the following as general guidelines to properly design equipment placement: 

•  

Place the hottest equipment (for example, the power supply), at the top of the rack.  

•  

Place precision equipment (such as the Model 260x) as low as possible in the rack (where temperatures 

 are 

cooler). 

 

•  

Add space panels below the Model 260x to help ensure adequate air flow. 

 

 

PA-909 Rev. C / May 2007

 

Summary of Contents for 4299-2

Page 1: ...e parts supplied with the Model 4299 2 Dual Unit Adjustable Heavy Duty Rack Mount Kit 4299 2 part listing Quantity Description Keithley part number 2 2 2 2 1 1 8 1 4 4 4 8 2 2 2 2 2 Hinge Front Rack E...

Page 2: ...low ensures proper cooling Adequate airflow provides air temperatures that are within about one inch of the Model 260x surfaces to remain within specified limits under all operating conditions DO NOT...

Page 3: ...ottom surface of the case and back until the orientation arrows on the handles line up with the orientation arrows on the mounting ears refer to Figure 1 With the arrows lined up pull the ends of the...

Page 4: ...ar Rack Mount Bracket 2602 334 to each instrument on the selected side using two Pan Head Sems Screws 10 32x3 8PPHSEM refer to Figure 2 Figure 2 Attach rear rack mount bracket Pan Head Sems Screw Phil...

Page 5: ...nstall the Rear Bracket 2304 327 securing it and the applicable rear bezels with the 6 32x5 8PPH screws and washers supplied Figure 3 Secure instrument together RA NG E AM PS 350 V PE AK AU TO SH IFT...

Page 6: ...ar rack rails and secure using two Truss Head Screws 10 32x5 8PHTRSH 3 Repeat for other side 4 Install Cage Nuts FA 274 into front rack rail holes aligned with the rear cage nuts refer to Figure 4 Fig...

Page 7: ...ity to the equipment and within easy reach of the operator 1 Slide units into rack by aligning both Rear Rack Mount Brackets 2602 334 with the associated Rack Mount Extension Arms 2602 335 refer to Fi...

Page 8: ...Model 4299 2 Installation Instructions Truss Head Screw Phllips 10 32x5 8PHTRSH 4 Places Front Rack Rail 2 places 8 PA 909 Rev C May 2007...

Reviews: