Section 6: Characterization of power discretes
Model 2651A High Power System SourceMeter® Instrument User's Manual
Figure 19: Trigger model configuration for IV curve characterization
Idle
Arm layer
Trigger layer
smua.trigger.
arm.count = 5
count = 21
IDLE_EVENT_ID
SWEEPING_EVENT_ID
SWEEP_COMPLETE_EVENT_ID
ARMED_EVENT_ID
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
arm.stimulus
source.stimulus
measure.stimulus
endpulse.stimulus
trigger.timer[1].
EVENT_ID
passthrough = true
count = 20
stimulus
tsplink.trigger[2].
EVENT_ID
stimulus
mode = tsplink.TRIG_FALLING
tsplink.trigger[1].
EVENT_ID
stimulus
mode = tsplink.TRIG_SYNCHRONOUSM
tsplink.trigger[2].
EVENT_ID
stimulus
mode = tsplink.TRIG_FALLING
tsplink.trigger[1].
EVENT_ID
stimulus
mode = tsplink.TRIG_SYNCHRONOUSA
node[1].
Model 2651A (master)
node[2].
Series 26xxA (subordinate)*
Idle
Arm layer
Trigger layer
smua.trigger.
arm.count = 1
count = 5
IDLE_EVENT_ID
SWEEPING_EVENT_ID
SWEEP_COMPLETE_EVENT_ID
ARMED_EVENT_ID
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
arm.stimulus
source.stimulus
measure.stimulus
endpulse.stimulus
trigger.timer[2].
EVENT_ID
passthrough = false
count = 1
stimulus
0.001 s
0.03 s
* Model 260xA, Model 261xA, or Model 263xA
6-6
2651A-900-01 Rev. A / March 2011