Quick Reference Guide
3
Basic connections
WARNING
A hazardous voltage condition exists at or above 42V peak. To prevent
electric shock that could result in personal injury or death, never make
or break connections while hazardous voltage is present.
CAUTION
Maximum Channel 1 voltage measurement is 120V peak. Maximum
Channel 2 measurement is 12V peak.
Optimizing connections
To ensure measurement integrity, observe the following when making connections:
•
Use copper-to-copper connections when possible to minimize thermal EMFs.
•
Any solder connections should use silver solder to minimize thermal EMFs.
•
Clean all connector terminals free of oxidation using a small amount of DeoxIT.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Pin 1
1
2
3
4
5
6
7
8
EXTERNAL TRIGGER INPUT
Trigger Reading
>2
µ
sec
TTL HI
TTL LO
Reading
Complete
VOLTMETER COMPLETE OUTPUT
>10
µ
sec
TTL HI
TTL LO
DIGITAL COMMON
Pins 7 and 8
RS232
120
Pin 2
1
3
5
2
4
6
VMC
EXT TRIG
FUSE
LINE
250mAT
(SB)
100 VAC
120 VAC
125mAT
(SB)
220 VAC
240 VAC
WITH ANALOG OUTPUT
GAIN SET TO 1:
±
FULL SCALE READINGS
PRODUCE
±
1V OUTPUT
TRIGGER
LINK
!
LINE RATING
50, 60
400HZ
17 VA MAX
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
!
!
ANALOG OUTPUT
1K
Ω
OUTPUT
RESISTANCE
MADE IN
U.S.A.
Figure 2
Rear panel