10
Quick Reference Guide
Programming examples
Voltage measurement example
— Table 4 shows a typical command sequence to
measure voltage on channel 1 using auto range.
Temperature measurement example
— Table 5 shows a typical command sequence to
measure temperature on channel 1 using a simulated junction with 0°C reference
temperature and a type J thermocouple.
Settings to optimize performance
Range
To achieve the best accuracy, the Model 2182 should be on the lowest possible
measurement range. In most situations, auto range can be used to automatically select the
best range. Auto range is controlled (enabled/disabled) by the AUTO range key (AUTO
annunciator indicates auto range is enabled).
Table 4
Command sequence for voltage measurement example
Command
1
Comments
*RST
Restore GPIB defaults.
:SENS:FUNC ‘VOLT’
Select voltage measurement function.
:SENS:CHAN 1
Select channel 1.
:SENS:VOLT:CHAN1:RANG:AUTO ON
Enable channel 1 auto range.
:READ?
Trigger and acquire one voltage reading.
1. Model 2182 must be addressed to talk after sending :READ? to trigger and acquire data.
Table 5
Command sequence for temperature measurement example
Command
1
Comments
*RST
Restore GPIB defaults.
:SENS:FUNC ‘TEMP’
Select temperature measurement function.
:SENS:CHAN 1
Select channel 1 for measurement.
:SENS:TEMP:TRAN TC
Select thermocouple sensor.
:SENS:TEMP:RJUN:RSEL SIM
Choose simulated junction.
:SENS:TEMP:RJUN:SIM 0
Set simulated junction temperature to 0˚C.
:SENS:TEMP:TC J
Select type J thermocouple.
:UNIT:TEMP F
Select ˚F temperature units.
:READ?
Trigger and acquire one temperature reading.
1. Model 2182 must be addressed to talk after sending :READ? to trigger and acquire data.