Quick Reference Guide
15
•
The STD DEV value is the standard deviation of the buffered readings. The equation
used to calculate the standard deviation is:
where: Xi is a stored reading.
n is the number of stored readings.
NOTE
The Model 2182 uses IEEE-754 floating point format for math calculations.
Remote buffer programming
SCPI commands
SCPI commands to configure and control the buffer and buffer statistics are listed in Table 7.
Table 7
Buffer SCPI commands
Command
Description
:TRACe:CLEar
Clear readings from buffer.
:TRACe FREE?
Query bytes available and bytes in use.
:TRACe:POINts <n>
Specify number of readings to store: 2 to 1024.
:TRACe:FEED <name>
Select source of readings: SENSe, CALCulate, or
NONE.
:TRACe:FEED:CONTrol <name>
Select buffer control mode: NEVer or NEXT.
:TRACe:DATA?
Read all readings in buffer.
:CALCulate2:FORMat <name>
Select buffer statistic: MINimum, MAXimum,
MEAN, SDEViation, or NONE.
:CALCulate2:STATe <b>
Enable or disable statistic calculation
:CALCulate2:IMMediate
Recalculate raw input data in buffer.
:CALCulate2:IMMediate?
Perform calculation and read result.
:CALCulate2:DATA?
Read result of statistic calculation.
X
2
i
i
1
=
n
∑
1
n
---
X
i
i
1
=
n
∑
2
–
n-1
--------------------------------------------------------------
y =