T-506 User Manual Version 3.2 | 23
5.6.2
Expert Mode (1) Key function
Button
Function
Mute or unmute the APP sound.
Pause or resume the automatic update of the sound and
magnetic field signal waveforms.
Store the current sound signal waveform.
Compare the stored sound signal waveform with the
currently displayed waveform (Clicking this button again
cancels the comparison function).
Select the filter mode of the sound channel. The submenu
functions are as follows:
‘Low pass’ band set to 100Hz
-
400Hz ‘Bandpass’ band is set
to 150Hz-600Hz
‘High Pass’ band is set to 200Hz
- 1500Hz
‘All
-
pass’ band is set to 100Hz
-1500Hz
(The default when the unit is powered on is
‘All
-
pass’ state)
.
Clicking
‘Confirm’
will set the selected filter mode and
‘Cancel’ will exit the Filter Settings submenu
This button will bring up the remote assistance interface to
initiate remote help.
Menu allowing language selection, software upgrades,
menu bar and About Us.
Switch to Operator mode
5.6.2.1
Information
This message will appear to provide confirmation of current test status and the identification
of the fault-discharge acoustic signal by the T-506 intelligent algorithm (e.g., magnetic field
trigger, fault found).