CALIBRATION 

Your RAPT Pill comes pre-calibrated from the factory. Due to minor tolerance differences, this 
calibration can be improved by performing a manual user calibration. 
 

1.

 

Connect your Pill via a USB-C cable to your Mac or PC, or connect your Mac or PC to your 

Pill’s Wifi network. Either will open the RAPT Pill Portal automatically

2.

 

Once the RAPT Pill Portal is open on your browser and your RAPT Pill is connected to your 
Wifi network (See 

How to connect a RAPT Pill to your Wifi

 Above), then access the IP 

address of the RAPT Pill by clicking on the network that the Pill is connected to in the Wifi 
screen. 

3.

 

Make a note of the IP address (usually in the form 192.168.0.XXXX). 

4.

 

Enter the IP address into your browser (we recommend Chrome). This will open the RAPT Pill 
Portal white the Pill is plugged into your Mac or PC with a USB-C cable. 

5.

 

Right click the calibration button and select ‘Open in New Tab’. Then do the same 

with the 

diagnostic button to open the Diagnostic screen in a new tab. 

IMPORTANT: The RAPT Pill is designed to disconnect from the RAPT Portal once the USB-C cable is 
unplugged to preserve battery life. Beginning the Calibration process with prevent this from 
occurring. Hence, keeping both the Calibration and Diagnostic screens open allows you to easily 
calibrate the RAPT Pill without losing connection to the Portal.

  

 
Having the Diagnostic screen open allows you to easily determine angle, temperature etc. via its real 
time telemetry which are required for advanced and custom calibration. 
 
For advanced and custom calibration, it is best to have Calibration open in one tab, and Diagnostics 
open in another. Note that you may be prompted to press a button to confirm that you wish to keep 
the portal open 

 this is normal. 

Once the diagnostic and Calibration screens are open, please assemble your Pill according to the 
directions so that it can be removed from the USB and calibrated according to the instructions 
below. 

There are three ways to calibrate your RAPT Pill. 

1: Standard Calibration 

Standard calibration is the quickest and simplest method of calibration. It is also potentially the least 
accurate. Fill a container of sufficient volume to hold it in suspension with a liquid of known gravity. 
It is simplest and easiest to use water. Once the angle reading has stabilised* enter the gravity 
(1.000 for water) and hit Calibrate. Wait until the screen indicates that calibration is complete. 

Your Pill is now ready to use! 

* The angle will never be completely steady, as it is extremely sensitive. Hit Calibrate when the RAPT 
Pill has stabilised as much as possible.  

Summary of Contents for RAPT KL20596

Page 1: ...RAPT Pill Hydrometer Thermometer Quick Start Guide Page 1 of 10 www KegLand com au Last Updated 1 12 2021 12 53 PM Pill Hydrometer Thermometer KL20596 Quick Start Guide KegLand Distribution PTY LTD...

Page 2: ...ts the battery from completing a connection and draining during shipping Battery Installation If required Only use KegLand approved 18650 rechargeable Lithium ion batteries KL26734 The RAPT Pill Hydro...

Page 3: ...lid Red Charging No Light Fully Charged ESP LED Flashing Green sending or receiving data telemetry SIGN UP FOR A RAPT ACCOUNT To utilise the Wi Fi connectivity of the RAPT Pill Hydrometer you will nee...

Page 4: ...ically open a browser to the Pill Captive Portal 3 Select your Wifi access point enter your password and select Join 4 Once you have joined click on the Registration link to see your MAC address Write...

Page 5: ...using it is important to insert the circuit board in the correct orientation to ensure the correct weight distribution for accurate readings The yellow end of the circuit board should be inserted into...

Page 6: ...STRUCTIONS Your RAPT Pill Hydrometer comes pre calibrated and ready to go after charging assembly and device registration For calibration steps refer to the Calibration section below 1 Submerge your R...

Page 7: ...ng the Calibration process with prevent this from occurring Hence keeping both the Calibration and Diagnostic screens open allows you to easily calibrate the RAPT Pill without losing connection to the...

Page 8: ...t does not matter if the gravity readings are not exactly 10 points apart as long as the gravity reading is correct Use of a hydrometer that is known to be accurate is quite important To perform the r...

Page 9: ...this y 0 0000023109x5 0 0005741632x4 0 0568746764x3 2 8119846525x2 71 1087310405x 265 6578215178 This is a sample only DO NOT USE THIS EQUATION AS IT WILL NOT BE ACCURATE Substitute x in this equatio...

Page 10: ...or support on how to use your RAPT Pill Hydrometer Thermometer or the RAPT Portal If you need hardware support please contact your nearest distributor The RAPT Pill Hydrometer Thermometer comes with a...