background image

 12

EN

KSL-Z by KEELER

6.  INSTRUCTIONS FOR USE

6.1  SETTING THE BINOCULARS

It is vital that the binoculars are optimised for the user’s optical 

correction in order to obtain focused binocular images.

1.  Remove the Test Bar locating hole cover plate (16) and place the test bar focus in the test 

bar location hole at the base of the microscope arm. To access the location hole fi rst remove 

the cover. The test bar should be set with the fl at projection face towards the Slit Lamp 

Microscope. The illumination and the microscope should be in the zero degrees position.

2.  Turn on the Slit Lamp and set the slit to full width (14), set the magnifi cation to x16 (27).
3.  Adjust the eyepieces pupillary distance by holding both eyepiece bodies and rotating 

them inwards or outwards until they are correct for your PD.

4.  Turn both eyepieces (28) to maximum plus (+) 

correction.

5.  Close one eye, and with the other eye look through 

the microscope slowly turning the open eye 

eyepiece towards the minus (-) position until the 

image of the test bar is in focus. Stop.

6.  Repeat the above process for the other eyepiece.
7.  Make a note of the positions of the eyepieces so that you can set them quickly if the Slit 

Lamp has been used by another clinician.

8.  Note – younger examiners are recommended to compensate for their ability to 

accommodate by further adjusting the eyepieces by minus one (-1) or minus two (-2) 

dioptres.

16

 

27

28

Summary of Contents for Z Series

Page 1: ...SLIT LAMP INSTRUCTIONS FOR USE Z Series A world without vision loss...

Page 2: ......

Page 3: ...OR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYSTEMS 15 7 2...

Page 4: ...ty Humidity limitation Use by date Serial number Catalogue number Medical device Translation The Keeler Slit Lamp is designed and built in conformity with Directive 93 42 EEC Regulation EU 2017 745 an...

Page 5: ...sembly and a table top with power supply and accessory drawer The light intensity is controlled by a variable rheostat located on the XYZ translation base There are a number of selectable filters allo...

Page 6: ...y the specialized technicians of the manufacturer s Technical Service Centre or by personnel trained and authorised by the manufacturer The manufacturer declines any and all responsibility for loss an...

Page 7: ...ortant when the unit has been stored or transported in a cold environment this can cause severe condensation to develop on the optical elements 2 3 CONTRAINDICATION There is no restriction to patient...

Page 8: ...on light 5 Chinrest 6 Chinrest height adjuster 7 Patient grab handles 5 3 4 1 2 7 6 13 14 15 8 16 17 18 19 11 10 12 9 KSL Z Series 8 Eyepiece assembly securing knob 9 Yellow filter knob up out 10 Illu...

Page 9: ...Runner covers 25 Z type tonometer mounting hole 26 Yellow filter knob up out 27 Lock for securing magnification body 28 Magnification change drum 29 Eyepieces adjustable for PD and dioptre correction...

Page 10: ...s castors ensure the following before moving it to another location 1 The table is at its lowest position 2 The power cord is removed 3 The Slit Lamp arm and base locking knobs are tightened 4 The run...

Page 11: ...Lamp base with the base notch b and pin c aligned Attach the tower to the base using the hex bolt removed earlier and tighten using the wrench provided 2 Carefully fit the microscope body to the arm e...

Page 12: ...le power supply see section 9 4 Power Supply page 20 2 If the Keeler Slit Lamp is used with a power supply or cables other than those supplied this may result in increased emissions or decreased immun...

Page 13: ...in use the Tonometer should be removed from the Slit Lamp and placed securely back in the packaging or a suitable location Applanation Tonometer Keeler Fixed Z TYPE This instrument is for those who wi...

Page 14: ...he slit to full width 14 set the magnification to x16 27 3 Adjust the eyepieces pupillary distance by holding both eyepiece bodies and rotating them inwards or outwards until they are correct for your...

Page 15: ...istance by holding both eyepiece bodies and rotating them inwards or outwards until they are correct for your PD 3 Switch on the illumination making sure the rheostat 21 is set to a low level to minim...

Page 16: ...e 0 parallel eyepieces 38mm 85mm Convergent angle of optical axis 13 Magnification change 5 step drum 3 step drum Magnification Field of view 3 5 6x 34mm 3 3 10x 22mm 3 3 16x 14mm 3 3 25x 8 5mm 3 5 40...

Page 17: ...with a soft clean lens cloth 2 Care must be taken to keep the objective and the eyepiece lenses clean use only soft clean lens cloths to clean optical surfaces 7 4 ELECTRICAL CONNECTIONS 1 Routinely c...

Page 18: ...g or using this instrument please read carefully and observe what is described here Portable or mobile type radio frequency communication units may have an adverse effect on this instrument resulting...

Page 19: ...to line s 2 kV line s for input output line s 1 kV line s to line s 2 kV line s for input output line s Mains power quality should be that of a typical professional healthcare facility Voltage dips sh...

Page 20: ...objects and people 1 Field strengths from fixed transmitters such as base stations cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot b...

Page 21: ...distance for the higher frequency applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 9 4 T...

Page 22: ...000 6 2 EN IEC 61000 6 3 Fuse T2 5AH 250V Power supply output 12V DC 2 5 amps must be EN IEC 60601 compliant Complies with Electrical Safety Medical EN IEC 60601 1 Electromagnetic compatibility EN IEC...

Page 23: ...e and transportation Do not store or transport this instrument in conditions where the temperature may rise above 50 C 10 ACCESSORIES AND SPARES Item Part Number Description Kapture Imaging Software L...

Page 24: ...ates that this product shall not be treated as household waste To reduce the environmental impact of WEEE Waste Electrical Electronic Equipment and minimise the volume of WEEE entering landfills we en...

Page 25: ......

Page 26: ...2021 USA Sales Office Keeler USA 3222 Phoenixville Pike Building 50 Malvern PA 19355 USA Toll Free 1 800 523 5620 Tel 1 610 353 4350 Fax 1 610 353 7814 India Office Keeler India Halma India Pvt Ltd P...

Reviews: