5.3 CABLE ATTACHMENT PROCEDURE
1. Connect the power cables.
a) Chinrest fi xation light cable to power supply unit.
b) (3 pin) cable from power supply unit to slit lamp base assembly.
c) (4 pin) main lamp cable from bottom of the illumination assembly to the slit lamp
base assembly.
d) Ensure cables are routed to allow free movement of the XYZ base and to be clear of
the patients.
If your Slit Lamp was not supplied with a transformer (Part #3020-P-5040),
make sure that the power connection is compatible with the specifi cations
in this manual and is connected by a qualifi ed technician to an available
and suitable power supply, see section 9.4 Power Supply page 20.
2. If the Keeler Slit Lamp is used with a power supply or cables other than those supplied,
this may result in increased emissions or decreased immunity of the Keeler Slit Lamp in
relation to EMC performance.
3. Connect the mains power to the Slit Lamp transformer using the power lead provided.
4. To isolate from the mains supply unplug the mains plug from the wall socket. Ensure
that the product is positioned so that it is easily accessible.
Only a hospital grade 3-conductor electrical power supply cable must be
used. For USA and Canada: Detachable power supply cord set, UL listed,
type SJE, SJT or SJO, 3-conductor, not smaller than 18 AWG. Plug, cable
and ground lead connection of the socket have to be in perfect condition.
4 pin
3 pin
3 pin to 3 pin
Fixation light socket