10.2 illumination toWer aSSembly procedure
1
Remove the hex bolt (a) from the base of the Illumination tower, and
then place the illumination tower on the Slit Lamp base with the base
notch (b) and pin(c) aligned. Attach the tower to the base using the hex
bolt removed earlier and tighten using the wrench provided.
2
Carefully fit the microscope body to the arm – ensuring it is pushed to
the stop. Tighten using the securing knob on the side.
3
Attach the breath shield to the pin on the rear of the magnification
assembly.
10.3 cable attachment procedure
1
Connect the main lamp cable from the Chinrest to the
illumination tower. Do not twist the lead on the tower
illumination system.
2
Connect the power cables
a) Chinrest fixation light cable to power supply unit
b) (3 pin) cable from power supply unit to slit lamp base assembly
c) (4 pin) main lamp cable from bottom of the chinrest to the slit lamp
base assembly.
d) Ensure cables are routed to allow free movement of the XYZ base and
to be clear of the patients.
If your Slit Lamp was not supplied with a transformer (Part #3020-
P-5040, make sure that the power connection is compatible with the
specifications in this manual and is connected by a qualified technician to
an available and suitable power supply,
see section 15.5 Power Supply
a
b
c
slide to stop and secure
4 pin
Cable connector
3 pin
Breath shield