7 cleaning and diSinfection inStructionS
Before any cleaning of the instrument or the base unit,
ensure the power lead is disconnected.
Only manual non-immersion cleaning as described should be used for
this instrument. Do not autoclave or immerse in cleaning fluids. Always
disconnect power supply from source before cleaning.
a Wipe the external surface with a clean absorbent, non-shedding cloth
dampened with a water / detergent solution (2% detergent by volume)
or water / isopropyl alcohol solution (70% IPA by volume). Avoid optical
surfaces.
b Ensure that excess solution does not enter the instrument.
Use caution to ensure cloth is not saturated with solution.
c Surfaces must be carefully hand-dried using a clean non-
shedding cloth.
d Safely dispose of used cleaning materials.
8 tranSport, Storage and Working conditionS
The following ambient condition limits are recommended for the Keeler
Slit Lamp, for transport and storage it is recommended that the Slit Lamp
is kept in its original manufacturers packaging.
Working environment
+10°C to +35°C
30% to 75% relative humidity
tranSport and Storage conditionS
Transport: -40°C to +70°C
Storage: -10°C to +55°C
Before use, the Slit Lamp should be allowed to adjust to
the ambient room temperature for several hours. This
is especially important when the unit has been stored or
transported in a cold environment; this can cause severe
condensation to develop on the optical elements.