Pulsair Intellipuff Tonometer by KEELER
• To prevent condensation from forming, allow instrument to come to room temperature
before use.
• Only mount on wall according to Keeler Instructions.
• This product should be used in a room with low / dimmed lighting.
• Before using the Pulsair IntelliPuff Tonometer, press the Demo button for 1 second
to dispel any minute particles of dust or moisture which may have settled whilst the
instrument was not in use.
• For indoor use only (protect from moisture).
• There are no user serviceable parts inside. Contact authorised service representative for
further information.
• Follow guidance on cleaning / routine maintenance to prevent personal injury / damage to
equipment.
• Failure to carry out recommended routine maintenance as per the instructions in this IFU
may reduce the operational lifetime of the product.
• At product end of life dispose of in accordance with local environmental guidelines (WEEE).
2.3 CONTRAINDICATION
There is no restriction to patient population this device can be used with other than those
outlined in the contraindications stated below.
Accuracy of IOP measurements is known to be affected by variations and changes in corneal
rigidity due to differences in corneal thickness, intrinsic structural factors or corneal refractive
surgery. It is recommended that these factors are considered during IOP measurement.
3. CLEANING INSTRUCTIONS
3.1 CLEAN THE PUFF TUBE LENS ON A WEEKLY BASIS:
1. Moisten a cotton bud with Isopropyl Alcohol.
2. Move the tip of the bud around the lens in a
circular motion.
3. After one circle the bud should be discarded
to avoid smearing on the lens.
4. Look at the Puff Tube lens from the patient’s
side, if traces of tear film can still be seen,
repeat above steps until clear.
Note:
Care should be taken not to damage the Puff
Tube assembly during cleaning.
CAUTION: Never use a dry cotton bud or tissue to clean the Puff Tube lens.
Never use a silicone impregnated cloth or tissue to clean the Puff Tube
lens.
Summary of Contents for Pulsair Intellipuf
Page 1: ...Non Contact Tonometer INSTRUCTIONS FOR USE EN BG RU HY HR SR...
Page 2: ......
Page 4: ......
Page 30: ...26 BG Pulsair Intellipuff KEELER Keeler Pulsair IntelliPuff Demo 1 2 3 3 3 1 1 2 3 4...
Page 33: ...29 BG Pulsair Intellipuff KEELER 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Page 72: ...68 HY KEELER Pulsair Intellipuff Keeler Pulsair IntelliPuff 1 Demo WEEE 2 3 3 3 1 1 2 3 4...
Page 75: ...71 HY KEELER Pulsair Intellipuff 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Page 92: ...88 KEELER Pulsair IntelliPuff RU Keeler Keeler EP29 32777 Pulsair IntelliPuff Keeler...
Page 93: ...89 KEELER Pulsair IntelliPuff RU Keeler Pulsair IntelliPuff Demo 1 WEEE 2 3 3 3 1 1 2 3 4...
Page 95: ...91 KEELER Pulsair IntelliPuff RU Pulsair Desktop 1 7 1 On Off 2 3 Print...
Page 96: ...92 KEELER Pulsair IntelliPuff RU 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Page 131: ......