17
EN
Pulsair Intellipuff Tonometer by KEELER
16.1 ELECTROMAGNETIC EMISSIONS
Guidance and manufacturer’s declaration – electromagnetic emissions
The Keeler Pulsair IntelliPuff Tonometer is intended for use in the electromagnetic environment
specified below. The customer or user should assure that it is used in such an environment.
Emissions test
Compliance Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The Keeler Pulsair IntelliPuff Tonometer uses RF energy
only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11
Class A
The Keeler Pulsair IntelliPuff Tonometer is suitable for use
in all establishments including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Complies
16.2 ELECTROMAGNETIC IMMUNITY
Guidance and manufacturer’s declaration – electromagnetic immunity
The Keeler Pulsair IntelliPuff Tonometer is intended for use in the electromagnetic environment
specified below. The customer or user should assure that it is used in such an environment.
Immunity test IEC 55015 Test
level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD).
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete
or ceramic tile.
If floors are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst.
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge.
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Summary of Contents for Pulsair Intellipuf
Page 1: ...Non Contact Tonometer INSTRUCTIONS FOR USE EN BG RU HY HR SR...
Page 2: ......
Page 4: ......
Page 30: ...26 BG Pulsair Intellipuff KEELER Keeler Pulsair IntelliPuff Demo 1 2 3 3 3 1 1 2 3 4...
Page 33: ...29 BG Pulsair Intellipuff KEELER 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Page 72: ...68 HY KEELER Pulsair Intellipuff Keeler Pulsair IntelliPuff 1 Demo WEEE 2 3 3 3 1 1 2 3 4...
Page 75: ...71 HY KEELER Pulsair Intellipuff 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Page 92: ...88 KEELER Pulsair IntelliPuff RU Keeler Keeler EP29 32777 Pulsair IntelliPuff Keeler...
Page 93: ...89 KEELER Pulsair IntelliPuff RU Keeler Pulsair IntelliPuff Demo 1 WEEE 2 3 3 3 1 1 2 3 4...
Page 95: ...91 KEELER Pulsair IntelliPuff RU Pulsair Desktop 1 7 1 On Off 2 3 Print...
Page 96: ...92 KEELER Pulsair IntelliPuff RU 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Page 131: ......