Instructions for use SMARTtorque S609 C - 1.008.1644 SMARTtorque Mini S605 C - 1.008.1645 SMARTtorque
S608 C - 3.001.6560
7 Processing steps in accordance with ISO 17664 | 7.7 Storage
22 / 26
Sterilization parameters:
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilization processes:
▪ Sterilizer with triple pre-vacuum:
- at least 3 minutes at 135°C (275 °F)
– Drying time: 16 min.
▪ Sterilizer using the gravity method:
- at least 10 minutes at 135°C (275 °F)
- drying time 30 minutes or alternatively
- at least 60 minutes at 121°C (250 °F)
- Drying time: 15 min.
▶
Remove contra-angle handpieces and turbines immediately after the com-
pletion of the sterilization cycle from the sterilizer.
▶
Use in accordance with the manufacturer's Instructions for Use.
7.7 Storage
Processed products must be stored, protected from bacteria, to the extent pos-
sible, and dust, in a dry, dark, cool room.
Note
Comply with the expiration date of the sterilized items.